FiberTEK III
Document 164802 Iss 4
User Manual
Page 13
Fig. 6 Loopback Field Calibration (3-Jumper)
•
Use the soft keys
/
to select whether to test a
single cable in one direction or a duplex cable in both directions. (This
setting does not affect the Field calibration.)
3.
Use soft key
to start the Field Calibration
4.
Follow the instructions on screen
5.
After the Field Calibration is complete, an indication of the quality of the jumper(s)
used is shown:
Fig. 7 Field Calibration result
NOTE
The cables connected to the Tx Ports must not be removed following calibration.
If field calibration was not successful, implement the following measures:
•
Check correct connection of launch cords.
•
Check launch fiber connections for contamination. If necessary, clean connections
and check with a fiber test microscope (IDEAL order number 45-332).
•
Dirty launch fibers also contaminate the measuring adapter connections. Clean
the connections on the FiberTEK™ III measuring adapters with an optical cleaning
sponge.
•
Check launch cords with a fiber continuity tester (IDEAL order number: VFF5) for
continuity.
•
Replace suspected damaged launch cords.