PERFORMANCE VERIFICAITONS
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning
correctly. If the meter does not conform to any of the listed limits the calibration procedure must be
performed.
Performance Verification Preparation
1. Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization
should be reached after 30 minutes.
2. Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5 V
DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test.
3. Input the values listed in Table 1 for the Standard 61-362
Table 2 Performance Verification 61-362
Function Setting /Range
Input
Low Limit
High Limit
ACV 3.2V
3.000V @ 50Hz
2.936
3.064
ACV 3.2V
3.000V @ 500Hz
2.936
3.064
ACV 32
30.00 @ 50Hz
29.36
30.64
ACV 32
30.00 @ 500Hz
29.36
30.64
ACV 320
300V AC @ 50Hz
293.6
306.4
ACV 320
300V AC @ 500Hz
293.6
306.4
ACV 600
500V AC @ 50Hz
486
514
ACV 600
500V AC@ 60Hz
486
514
DCV 320mV
300mV DC
296.3
303.7
DCV 3.2V
3.000V DC
2.975
3.025
DCV 32V
30.00V DC
29.63
30.37
DCV 320
300.0V DC
296.3
303.7
DCV 600
500V DC
493
507
A DC 320µA
300µA 293.9
306.1
A DC 3200 µA
3000µA 2939
3061
A DC 32mA
30mA
29.39
30.61
A DC 320mA
300mA 293.9
306.1
A DC 10
9.00 DCA
8.70
9.30
A AC 320µA
300µA @ 50Hz
292.1
307.9
A AC 320µA
300µA @ 500Hz
292.1
307.9
A AC 3200µA
3000µA @ 50Hz
2921
3079
A AC 320oµA
3000µA @ 500Hz
2921
3079
A AC 32mA
30mA @ 50Hz
29.21
30.79
A AC 32mA
30mA @ 500Hz
29.21
30.79
A AC 320mA
300mA @ 50Hz
292.1
307.9
A AC 320mA
300mA @500Hz
292.1
307.9
A AC 10A
9A @ 50Hz
8.64
9.36
A AC 10A
9A @ 500Hz
8.64
9.36
Ω
320
300.0 297.7
306.3
Ω
2k
3.000k 2.950
3.048
Ω
32k
30.00k 29.52
30.48
Ω
320k
300.0k 295.2
304.8
Ω
2M
3.000M 2.922
3.078
Ω
32M
30.00M 28.45
31.55
Diode Test
500mV
.448
.552
Continuity Test
120
Ω
Beep off
Continuity Test
80
Ω
Beep on
Form number 61362
rev 1 May 2002