Form number TM61361
Rev 3 November 2002
Page 7
Replacing Fuse
1. Disconnect the test leads and turn the range switch to “OFF”.
2. Remove the three screws holding the bottom case cover
3 Use a digital multimeter in low resistance {ohms} mode to check the two fuses
MFD input, 0.1A / 250V fast acting fuse.
2A (R/L1) input, is a 2A / 600V fast acting fuse
R/L2 input, is not fused
4. Remove the defective fuse and replace with the recommended fuse
LA-3893 (2A / 600V) and LA-3898 (0.1A / 250V) types are recommended.
5. Install bottom case cover and secure with screws.