Page 5
Table 2 Performance Verification
61-361 True RMS
Function Setting /Range
Input
Low Limit
High Limit
ACV 200mV
190mV AC@ 50Hz
185.8
194.2
ACV 200mV
190mV AC@ 500Hz
185.8
194.2
ACV 2000mV
1900mV @ 50Hz
1862
1938
ACV 2000mV
1900mV @ 500Hz
1862
1938
ACV 200V
190V AC @ 50Hz
185.8
194.2
ACV 200V
190V AC @ 500Hz
185.8
194.2
ACV 600V
500V AC @ 50Hz
486
514
ACV 600V
500V AC@ 60Hz
486
514
DCV 200mV
190mV DC
188.9
191.1
DCV 2000mV
1900mV DC
1889
1911
DCV 20V
19.00V DC
18.89
19.11
DCV 200V
190.0V DC
188.9
191.1
DCV 600V
500V DC
496
504
ADC 2A
1.900 DCA
1.848
1.952
AAC 2A
1.900 AAC @ 50Hz
1.839
1.961
AAC 2A
1.900 AAC @ 500Hz
1.839
1.961
Ω
200
100.0 98.6
101.3
Ω
2K
1.000K .986
1.014
Ω
20K
10.00K 986
10.14
Ω
200K
100.0K 98.6
101.4
Ω
20M
10.00M 9.76
10.24
Ω
200M
100.0M 94.0
106.0
200 µF
100.0µF
95.0
105.0
2mF 1mF
.950
1.050
20mF
10mF
9.50
10.50
Diode Test
500mV DC
485
515
Continuity Test
50
Ω
Beep off
Continuity Test
20
Ω
Beep on
Hz 10KHz
@10V
9.77
10.23
Hz
100KHz @ 10V
97.7
102.3
Phase Rotation Check. (RST)
1
RST
R S T will be displayed in the upper left corner if all three supply lines are powered
If meter indicates an “OK” message in display phase rotation is Clockwise
If meter indicates An “ER” message in display phase rotation is Counter Clockwise
Note 1:
Three-phase rotation test can be performed on any known three-phase source.
Form number TM61361
Rev 3 November 2002