Form Number TM61312-4
Rev 2 Aug 2007
Page 16
Table 4a AC microampere Test:
Step
Range Source
Frequency Reading
1
400µA
40µA
60Hz
29.0 to 40.9
2
400µA
40µA
400Hz
29.0 to 40.9
3
400µA
300.0µA
60Hz
295.0 to 305.0
4
400µA
300.0µA
400Hz
295.0 to 305.0
5
4000µA
3000µA
60Hz
2950 to 3050
6
4000µA
3000µA
400Hz
2950 to 3050
Testing the AC milliampere function: (for 61-312 and 61-314)
1. Connect the calibrator to the
V
Ω
Hz
and
COM
inputs on the meter.
2. Turn the rotary switch to mA
.
a. Press
the
SEL
button to select the
AC
function
3. Apply the inputs for steps 1 through 4 in Table 4b.
4. For each input, compare the reading on the meter display to the reading for your m
in Table 4b
eter
t
5. If the display reading falls outside of the range shown in Table 4b, the meter does no
meet specification.
Table 4b AC mA Test:
Step
Range
Source Frequency
Reading
1
40mA
30.00mA
60Hz
29.50 to 30.50
2
40mA
30.00mA
400Hz
29.50 to 30.50
3
400mA
300.0mA
60Hz
295.0 to 305.0
4
400mA
300.0mA
400Hz
295.0 to 305.0
Testing the AC A function (for 61-312 / 61-314)
1. Connect the calibrator to the
V
Ω
Hz
and
COM
inputs on the meter.
2. Turn the rotary switch to A
.
a. Press
the
SEL
button to select the
AC
function
3. Apply the inputs for steps 1 through 4 in Table 4c.
4. For each input, compare the reading on the meter display to the reading for your m
in Table 4c
eter
t
5. If the display reading falls outside of the range shown in Table 4c, the meter does no
meet specification.