Form Number TM61312-4
Rev 2 Aug 2007
Page 14
Table 2 AC Voltage Test:
Step Range Input
Frequency
Reading 61-312
Reading 61-314
1
4V
3.000V
60Hz
2.959 to 3.041
2.959 to 3.041
2
4V
3.000V
400Hz
2.950 to 3.050
2.947 to 3.053
3
40V
30.00V
60Hz
29.59 to 30.41
29.59 to 30.50
4
40V
30.00V
400Hz
29.50 to 30.50
29.47 to 30.53
5
400V
300.0V
60Hz
295.9 to 304.1
295.9 to 304.1
6
400V
300.0V
400Hz
295.0 to 305.0
294.7 to 305.3
7
600V
600V
60Hz
588 to 612
588 to 612
8
600V
600V
400Hz
586 to 614
583 to 617
Testing the DC microamperes Function (for 61-312 and 61-314)
To verify the accuracy of
AC
and
DC
current measurement functions, do the following:
1. Connect the calibrator to the
V
Ω
Hz
and
COM
inputs on the meter.
2. Turn the rotary switch to µA .
a. Press
the
SEL
Button to select the
DC
function
3. Apply the inputs for steps 1 through 3 in Table 3a.
4. For each input, compare the reading on the meter display to the reading for your m
in Table 3a
eter
t
5. If the display reading falls outside of the range shown in Table 3a, the meter does no
meet specification.
Table 3a DC microamperes Test:
Step
Range Source
Reading
1
400µA
4.0µA
3.5 to 4.5
2
400µA
300.0µA
295.9 to 304.1
3
4000µA
3000µA
2959 to 3041
Testing the DC milliamperes Function (for 61-312 and 61-314)
1. Connect the calibrator to the
V
Ω
Hz
and
COM
inputs on the meter.
2. Turn the rotary switch to mA
.
a. Press
the
SEL
button to select the
DC
function
3. Apply the inputs for steps 1 through 2 in Table 3b.
4. For each input, compare the reading on the meter display to the reading for your m
in Table 3b
eter
t
5. If the display reading falls outside of the range shown in Table 3b, the meter does no
meet specification.