32
E28IB004EN-G
1. With the supplied Allen key, unscrew the screw fixing the battery
compartment to the rear of the instrument’s display (see fig. 11).
2. Lift off the battery compartment.
The screw stays loosely attached to the battery compartment so that it
cannot get lost.
3. Remove the cover of the battery compartment by pressing on it and
sliding it forward simultaneously (see fig. 12).
Fig. 12: Removing the cover of the battery compartment
4. Exchange the batteries for new ones.
5. Reassemble all parts in reverse order.
11.6 Exchanging the Pump Lever
1. Dismount the pump as described in section 11.2.
2. The pump lever sits on the two pivot pins. Pull it off carefully by hand.
Do not use any tools.
3. Mount the new pump lever by shifting it in place.
4. Remount the pump as described in section 11.2.
NOTICE
• Observe the correct polarity when you insert the batteries, see the
engraving in the battery compartment.
• Use only batteries of the same type and with the same charge level.
• Ensure that the interior of the instrument is perfectly dry. Any kind of
liquid will cause damage to the electronic parts of the instrument.