WF-2000 Series AIO
User’s Manual (Ver. 1.1, Mar./2014) ------------- 46
Type Code
Input Range
Data Format
+F.S
-F.S
Engineering Unit
+1
-1
0Ah
-1
~
+1 V
2’s comp HEX
7FFFh
8000h
Engineering Unit
+500
-500
0Bh
-500
~
+500 mV
2’s comp HEX
7FFFh
8000h
Engineering Unit
+150
-150
0Ch
-150
~
+150 mV
2’s comp HEX
7FFFh
8000h
Engineering Unit
+20
-20
0Dh
-20
~
+20 mA
2’s comp HEX
7FFFh
8000h
Engineering Unit
+760
-210
0Eh
Type J
Thermocouple
-210
~
760°C
2’s comp HEX
7FFFh
DCA2h
Engineering Unit
+1372
-270
0Fh
Type K
Thermocouple
-270
~
1372°C
2’s comp HEX
7FFFh
E6D0h
Engineering Unit
+400
-270
10h
Type T
Thermocouple
-270
~
400°C
2’s comp HEX
7FFFh
A99Ah
Engineering Unit
1000
-270
11h
Type E
Thermocouple
-270
~
1000°C
2’s comp HEX
7FFFh
DD71h
Engineering Unit
+1768
0
12h
Type R
Thermocouple
0
~
1768°C
2’s comp HEX
7FFFh
0000h
Engineering Unit
+1768
0
13h
Type S
Thermocouple
0
~
1768°C
2’s comp HEX
7FFFh
0000h
Engineering Unit
+1820
0
14h
Type B
Thermocouple
0
~
1820°C
2’s comp HEX
7FFFh
0000h
15h
Type N
Engineering Unit
+1300
-270