1.6.7 Set Operation Mode to PulseDir Mode
step command
response
1 $01M
!017083B
2 $01D02
!01
3 $01D16
!01
4 $01D2E
!01
5 $01S0
!0120
6 $01S1
!0160
7 $01S2
!01E0
8 $01S3
!0100
z
step 2: set encoder 0
Æ
PulseDir mode
XOR0=0, input signal is single-ended & active high (Sec. 1.5.2)
L0=0, the preset value no update when power if turn OFF
z
step 3: set encoder 1
Æ
PulseDir mode
XOR1=1, input signal is differential (Sec. 1.5.1)
L1=0, the preset value no update when power is turn OFF
z
step 4: set encoder 2
Æ
PulseDir mode
XOR2=1, input signal is differential (Sec. 1.5.1)
L2=1, encoder value will save to the preset value when power is
turn OFF. It is the default setting of 7083B(D)
z
step 5: mode read back=PulseDir, XOR0=0, L0=0,
Z0=A0=B0=Low
Æ
Z,A,B must be normal Low & active High.
z
step 6: mode read back=PulseDir, XOR1=1, L1=0,
Z1=A1=B1=Low
Æ
Z,A,B must be normal Low & active High.
z
step 7: mode read back=PulseDir, XOR2=0, L2=1,
Z2=A2=B2=Low
Æ
Z,A,B must be normal Low & active High
z
step 8: read all Z, Z2=Z1=Z0=Low
I-7083/7083D/7083B/7083BD User Manual (V 1.1, Dec/2007) ------------------ 19