D500 | D520 Series Operating Manual

25

◀◀

|

The following illustrates the request frame and response frame with the read and hold register 108-110 as an example. (The contents of 

register 108 are read-only, with two-byte values of 0X022B, and the contents of register 109-110 are 0X0000 and 0X0064)

Request Frame

Response Frame

Number Systems

(Hexadecimal)

Function code

(Hexadecimal)

Function code

0x03

Register Value (High Bytes) (108)

0x03

Start address (high byte)

0x00

Register Value (Low Bytes)(108)

0x06

Start address (low byte)

0x6B

Register Value (High Bytes) (109)

0x02

Number of Read Registers (High Bytes)

0x00

Register Value (Low Bytes) (109)

0x2B

Number of Read Registers (Low Bytes)

0x00

Register Value (High Bytes)(110)

0x00

Register Value (High Bytes)(110)

0x00

Register Value (Low Bytes) (110)

0x00

Function code

0x64

Figure 10: Examples of read and hold register request and response frames

MODBUS function code 0x10: write multiple registers 

This function code is used to write continuous registers to remote devices (1... 123 registers) block that specifies the value of the registers 

written in the request data frame. Data is packaged in two bytes per register. Response frame return function code, start address and 

number of registers written.

Request:

Function Code

1 byte

0x03

Start Address

2 bytes

0x0000….0xfffff

Number of Input Registers

2 bytes

1...125

Number of Bytes

1 byte

Nx2

Register Values

N×2 bytes

Value

N = Register number 

Figure 11: Write multiple register request frames

Response:

Function Code

1 byte

0x03

Start Address

2 bytes

0x0000….0xfffff

Number of Input Registers

2 bytes

1...125

Number of Bytes

1 byte

Nx2

Register Values

N×2 bytes

Value

N = Register number 

Figure 12: Write multiple register response frames

Содержание proCon D500 Series

Страница 1: ...D500 D520 Series pH ORP Digital Analyzer Controller Operating Manual...

Страница 2: ...ration 12 pH Buffer Reading vs Temperature 13 Field Calibration 13 Data Logging Graphic Trending Trend Chart 14 Measurement Settings 16 SimplCal Calibration 17 Field Calibration 18 Calibration Adjusme...

Страница 3: ...tion or repair For example 1 Apparent damage to the analyzer 2 The analyzer does not work properly or provides specified measurements 3 The analyzer has been stored for a long time in an environment w...

Страница 4: ...54 F 0 3 C Temperature Compensation Manual or Automatic Stability pH 0 01pH 24h ORP 1mV 24h Outputs Two 2 4 20mA 20 4mA 0 20mA Communication RS485 MODBUS RTU Relay Control Contact 5A 240VAC 5A 28VDC P...

Страница 5: ...ion Output G TX RX RS232 Communication Output P P DC Power T2 T2 Temperature Connection EC1 EC2 EC3 EC4 Conductivity Resistivity Connection RLY3 RLY2 RLY1 Three Relays L N Communication Terminal L Liv...

Страница 6: ...500 D520 Series Operating Manual 6 Instrument Installation 1 Cable Recommended stripping length for cables is at least 90mm please use 0 5 to 1 square meter wire 2 Cable ties 3 Waterproof cable glands...

Страница 7: ...D500 D520 Series Operating Manual 7 Slim Line Wall Mount Bracket Typical Application...

Страница 8: ...ORP Sensor 1 pH ORP Sensor 2 Relay 1 Pump Valve Relay 2 Pump Valve Relay Alarm Cleaning 4 20mA 4 wire No Pre Amp Required No Pre Amp Required D400 R PH ORP 4 001pH 14 51 mA C1 SP 1 SP 2 SP 3 pH ORP 2...

Страница 9: ...021 11 09 12 53 17 pH ORP 14 51 mA pH Value Unit Year Month Day Sensor Communication Warning Hour Minutes Seconds Relay 1 blue is off Relay 2 red is on Relay 3 Sensor Cleaning Relay Instrument Type Cu...

Страница 10: ...IST 6 86 4 01 9 18 6 86 4 01 9 18 ORP Standard Solution 235mV Default Field Calibration Field Calibration Enter Value Offset Adjustment Enter Value Slope Adjustment Enter Value Calibration Adjustment...

Страница 11: ...1 Bit 2 BIT Network Node 000001 Data Log Graphic Trend Trend Chart Record Interval Display according to interval settings 480 points screen 1h point 12h point 24h point Record Query Data Points Record...

Страница 12: ...y password PH Calibration pH ORP 14 51 mA C1 SP1 SP2 SP3 7 00 4 01 10 01 5 mV 170 mV 5 mV 98 Done Calibrating Error Offset Slope 2023 11 09 12 53 17 2023 11 09 12 53 17 pH ORP 14 51 mA C1 SP1 SP2 SP3...

Страница 13: ...2 9 73 ORP Calibration Press the MENU button to enter the calibration mode select ORP standard liquid calibration and input the known standard liquid value Default 235mV Press the MENU button to retur...

Страница 14: ...unit Field Calibration 7 5s Store data every 7 5 seconds 90s 1 3 minutes Store data every 90 seconds 1 5 minutes 180s 3 minutes Store data every 180 seconds 3 minutes Press the MENU button returns to...

Страница 15: ...D500 D520 Series Operating Manual 15...

Страница 16: ...D500 D520 Series Operating Manual 16 Measurement Settings...

Страница 17: ...D500 D520 Series Operating Manual 17 SimplCal Calibration...

Страница 18: ...D500 D520 Series Operating Manual 18 Field Calibration...

Страница 19: ...D500 D520 Series Operating Manual 19 Calibration Adjusment...

Страница 20: ...D500 D520 Series Operating Manual 20 Alarm Relay Settings...

Страница 21: ...D500 D520 Series Operating Manual 21 Output Settings...

Страница 22: ...or example YL0114010022 Command Structure The MODBUS application protocol defines the Simple Protocol Data Unit PDU which is independent of the underlying communication layer Figure 1 MODBUS Protocol...

Страница 23: ...y Each character or byte is sent in this order from left to right the least significant bit LSB Maximum Significant Bit MSB Start bit 1 2 3 4 5 6 7 8 Stop bit Stop bit Figure 3 RTU pattern bit sequenc...

Страница 24: ...3 5 characters 1 247 Function codes conforming to MODBUS specifi cation Data conforming to MODBUS specification CRCL CRCL No signal bytes during 3 5 characters Byte 3 5 1 n 1 1 3 5 Figure 7 MODBUS de...

Страница 25: ...Value Low Bytes 110 0x00 Function code 0x64 Figure 10 Examples of read and hold register request and response framesMODBUS function code 0x10 write multiple registers This function code is used to wr...

Страница 26: ...number in binary form first finding the binary representation of the integer part 17 Decimal 16 1 1 24 0 23 0 22 0 21 1 20 The binary representation of integer part 17 is 10001B then the binary repres...

Страница 27: ...10 0111 1011 0110 0110 0110 0110B to decimal number Step 1 Divide the binary float ing point number 0100 0010 0111 1011 0110 0110 0110 0110B into symbol bit exponential bit and mantissa bit 0 10000100...

Страница 28: ...then shortcircuit the terminal Input and Ref the instrument should be displayed in 7 000 or within the error range If the display is not correct please contact the manufacturer for maintenance Please...

Страница 29: ...one year from the date of purchase If repair is necessary and has not been the result of abuse or misuse within the warranty period please return to Icon Process Controls and amendment will be made w...

Страница 30: ...30 22 0004 01 27 23 Need a great sensor Pair your D500 D520 Series controller with the industry s most innovative and reliable sensing technology iconprocon com...

Отзывы: