Electrical Wiring
Fig. 7: Pin code: Connection plug / Socket for 5-pin version
Serious injury or death from electric shock may occur if wiring, installation, disassembly or remove of wires is
performed while electrical power is energized.
Always shut off or disconnect electrical power at service panel and lock switch or breaker and tag to prevent energizing
electrical power during work or while Flowmeter is not assembled and installed.
Wiring installation, disassembly and removal must be performed by qualified persons experienced and knowledgeable
about electrical work.
1
24 VDC
Power supply : 18...30 VDC
5-Pin Plug with 1-Wire Communication
Pin
Function
Description
3
GND
4
Communication
Ground : 0 V
Communication interface
5
Analog output QA
4-20mA ; 0-20mA
Example: 0L/min => 4mA
300L/min => 20mA (depending on diameter)
Empty pipe Alert => 3.5mA
4
1
3
5
2
4
3
1
5
2
2
Pulse / Frequency output
Alternative:
Digital Output Q1
Freely adjustable ranging from 0.1 to 3000 ml/pulse in 0.1 ml/pulse steps,
npn-Transistor, max. load 30V/100mA.
Max. Voltage must be less than the supply voltage.
1. Empty pipe output
2. Dosing output
3. Upper or Lower Limit output
4. Negative flow
Configurable output of 0V or 24V when pipe is empty.
Configurable output of 0V or 24V
Configurable output of 0V or 24V when reaching upper or lower limit
Configurable output of 0V or 24V when liquid flows in negative direction
1
24 VDC
Power supply : 18...30 VDC
5-Pin Plug with RS485 Communication and Current Output
3
GND
4
Communication
Ground : 0 V
Communication interface
5
Analog output QA
4-20mA ; 0-20mA
Example: 0L/min => 4mA
300L/min => 20mA (depending on diameter)
Empty pipe Alert => 3.5mA
2
Communication
RS 485 B
www.iconprocon.com | [email protected] | 905.469.9283
08
Ultrasonic Flow Meter
UltraFlo
®
2000 SERIES