IceCure Medical Ltd.
DSR3000000 rev.L
Confidential
User Manual
53
IceSense3™
Warning
Liquid nitrogen may cause serious injury or burn if handled improperly. Local laws and safety rules
regarding liquid nitrogen Dewars should always be observed. Maintenance of liquid nitrogen
Dewars should be performed by authorized personnel only.
Check that the Dewar is fully filled (see the arrow in Fig. 40) and cover it with the sponge lid to
transfer it back to the system.
Check there is no frost on the Dewar after you filled it.
In the event that a Dewar is damaged, use another Dewar to continue the procedure.
Warning
Do not use a liquid nitrogen Dewar if it is damaged.
You can tell that a Dewar is damaged if after filling it, frost appears on the outer wall of the
container. Return the Dewar to IceCure technician or an authorized distributor for inspection
Warning
Do not transfer a Dewar with Liquid nitrogen unless it is covered with a lid.
Take off the lid and place the Dewar in its position into the system.
Warning
Removing the Dewar, or placing it back within the system after refilling it must ONLY be done
according to system instruction and with the carriage in the bottom position. If the carriage is not
in the bottom position, liquid nitrogen may spill out.
Close the compartment door and press the “Action” button on the cryohandle or “Next” on
the screen.
Dewar door is open
If the system detects that the Dewar compartment door is open when it tries to move the Dewar
carriage up or down, the following popup message will appear:
Dewar door is open. Close the door to continue.