![IceCure IceSense3 Скачать руководство пользователя страница 37](http://html2.mh-extra.com/html/icecure/icesense3/icesense3_user-manual_3374917037.webp)
IceCure Medical Ltd.
DSR3000000 rev.L
Confidential
User Manual
37
IceSense3™
Figure 23: Handle plug holder and handle plug.
3.3.7
Foot pedal (Not available in some regions, e.g. China)
The pedal is an alternative to the Action button on the cryohandle; pushing the pneumatic foot pedal
is identical to pushing the Action button.
Figure 24: The foot pedal
3.3.8
Temperature sensor (TS)
The temperature sensor is a single use accessory that provides real time temperature measurements
when inserted in tissue. Use of temperature sensors is optional; the decision to use a temperature
sensor must be made before the system pretest is initiated.
Prior to starting a procedure, the temperature sensor is connected to the system and inserted into
the desired location within the tissue. Insert the sensor at location that gives useful information
about the progress of the freezing procedure. For example, the temperature sensor can be placed
near the boundary of a tumor to ensure that the entire tumor has been frozen. Alternatively, the
temperature sensor may be placed near an organ that should not be frozen during the procedure to
ensure that the ice front has not advanced to this location.
The temperature sensor is not indicated for measuring the body temperature, but only for getting
temperature indication of some tissue during cryoablation