GUIDE
MEASUREMENT
1. Select the eye you want to measure using
the left and right navigation buttons.
2. Bring the tonometer near the patient’s
eye. **Note: the angle must be 90
°
in
both horizontal and vertical measure-
ment. **
3. The distance from the tip of the probe to
the cornea must be 3-7mm (1/8 - 2/7
inch). If necessary, adjust the distance
using the forehead support.
4. Press the main button lightly to perform one
individual measurement, taking care not to
shake the tonometer. The tip of the probe
should make contact with the central cornea.
5. Make six single measurements. Once the six
measurements are complete, the final IOP is
displayed. Save the measurement result by
pressing main button after results!
6. Please refer to the manual for
additional information.
ONE SHORT BEEP
–
Indicates a correct single measurement.
DOUBLE BEEP
––
Indicates an incorrect measurement.
LONG BEEP
Indicates that six succesfull measure-
ments are done and IOP result is available
at the display.
From the main menu, select “Turn off” and
confirm by pressing main button twice.
If not used for three minutes, the
tonometer will automatically switch off
3–7 mm
(1/8 - 2/7’’)
90
�
Correct measuring positionings with Icare
PRO are horizontal and vertical.