background image

Icare HOME Patient Guide

ENGLISH

15

www.icare-usa.com

14

Press the Measure button         to measure 

pressure (probe will contact eye during 

measurements). 

Press and hold the button (3 seconds) to collect 6 

measurements in a row, or press the button briefly (1 

second) 6 times to collect one measurement at a time.

 
     

PRECAUTION

If an error occurs (two short beeps and the probe 

base light is flashing red), press the Measure but-

ton briefly (1 second) and continue measurement.

When the measurement ends successfully, 

you will hear a long beep, the probe base 

light will power off, and the Done light 

illuminates. 

After using the tonometer to measure your eye 

pressure, you can remove the probe by powering 

off the device and allowing the probe to slide 

out of the probe base. Place the probe back in 

its container and dispose of it in a normal trash 

container or a metal waste bin.

     

WARNING

You should replace the probe if it comes into

contact with your eyelashes or eyelid or you 

suspect it did.

If your doctor instructed you to measure both 

eyes, repeat steps 1 through 10 using your 

other eye. Use a new, unused probe after mea-

surements for both eyes have been taken.

After completing all measurements, power 

off the tonometer by pressing and holding the 

Power button          down until you hear 3 short 

beeps.

 
     

PRECAUTION

The tonometer will power off automatically if 

unused for 3 minutes. 

COMPLETING THE 

MEASUREMENT

9

10

11

12

i

i

Содержание Icare HOME TA022

Страница 1: ...1 www icare usa com PATIENT ENGLISH Federal US law restricts this device to sale by or on the order of a physician or properly licensed practitioner Belongs to GUIDE...

Страница 2: ...clear layer Forehead cheek supports tonometer s adjustable supports Expected service life expected life time before replacement Electromagnetic immunity tonometer s ability to withstand disturbances...

Страница 3: ...have taken the measurements for both eyes The tonometer includes infra red eye recognition sensors to identify which eye right or left you are measuring It is important not to cover these sensors with...

Страница 4: ...oor or off center visual fixation Have poor hearing without an aid and or communicate using sign language Use contact lens Have dry eyes Have Keratoconus a corneal disorder Have a congenitally from bi...

Страница 5: ...indicator REPEAT indicator DONE indicator SERVICE indicator BATTERY indicator Power button Cheek support position indicator Device label Cheek support adjustment wheel Silicone lid USB cover Battery c...

Страница 6: ...es made by the manufacturer Do not touch the probe The shelf life of the probes is 3 years Use a new unused probe after measure ments for both eyes have been taken You should thoroughly wash your hand...

Страница 7: ...the probe tip points straight at cornea WARNING Make sure that the probe or any part of the tonometer does not contact your eyes PRECAUTION Do not cover the eye recognition sensors because covering th...

Страница 8: ...onometer to measure your eye pressure you can remove the probe by powering off the device and allowing the probe to slide out of the probe base Place the probe back in its container and dispose of it...

Страница 9: ...The probe or probe base is dirty or damaged Battery light is flashing Cause Battery power is low ERROR MESSAGE 1 Make sure you are facing straight ahead i e head held at a 90 angle 2 Adjust the tonom...

Страница 10: ...d batteries and insert 2 new CR123A lithium batteries Make sure the batteries positive end is facing the battery cover Rechargeable batteries should not be used Slide the battery cover back over the b...

Страница 11: ...damp cloth or paper towel Make sure the batteries positive end is facing the battery cover PRECAUTION Do not immerse the tonometer in water or any other liquid PRECAUTION Any maintenance services othe...

Страница 12: ...f 5 to 50 mmHg Compared to the Goldmann tonometer less than 5 of measurements fell outside 5 mmHg at each pressure range and less than 1 fell outside 7 5 mmHg at each pressure range The Icare HOME s v...

Страница 13: ...mbined with ISO 13485 certified quality system has made us a respected player in our field of expertise The advanced Icare product line offers reliable high pre cision reproducible accuracy in measuri...

Отзывы: