Icare HOME Patient Guide

ENGLISH

13

www.icare-usa.com

12

Red indicator light

Green indicator light

Green indicator light

Green indicator light

Stand in front of a mirror and hold the 

tonometer sideways in front of your face. 

This will help you to align the tonometer 

to your eye.

     

WARNING

Make sure that the probe or any part of the 

tonometer does not contact your eyes.

Align the probe tip with center of cornea

and rotate the tonometer until the

probe tip points straight at cornea.

     

WARNING

Make sure that the probe or any part of the 

tonometer does not contact your eyes.

     

PRECAUTION

Do not cover the eye recognition sensors, 

because covering the sensors causes an error.

Make sure the probe base light is green.

If the probe base light is red, make sure you 

are facing straight ahead (i.e., head held at 

a 90° angle) and tilt the tonometer until the 

probe base light turns green.

Make sure the probe is centered in sight

to ensure the probe contacts the center of

the cornea during measurement.

If the probe is not centered in your sight,

repeat steps 5 and 6.

POSITIONING THE TONOMETER

5

6

7

8

i

Содержание Icare HOME TA022

Страница 1: ...1 www icare usa com PATIENT ENGLISH Federal US law restricts this device to sale by or on the order of a physician or properly licensed practitioner Belongs to GUIDE...

Страница 2: ...clear layer Forehead cheek supports tonometer s adjustable supports Expected service life expected life time before replacement Electromagnetic immunity tonometer s ability to withstand disturbances...

Страница 3: ...have taken the measurements for both eyes The tonometer includes infra red eye recognition sensors to identify which eye right or left you are measuring It is important not to cover these sensors with...

Страница 4: ...oor or off center visual fixation Have poor hearing without an aid and or communicate using sign language Use contact lens Have dry eyes Have Keratoconus a corneal disorder Have a congenitally from bi...

Страница 5: ...indicator REPEAT indicator DONE indicator SERVICE indicator BATTERY indicator Power button Cheek support position indicator Device label Cheek support adjustment wheel Silicone lid USB cover Battery c...

Страница 6: ...es made by the manufacturer Do not touch the probe The shelf life of the probes is 3 years Use a new unused probe after measure ments for both eyes have been taken You should thoroughly wash your hand...

Страница 7: ...the probe tip points straight at cornea WARNING Make sure that the probe or any part of the tonometer does not contact your eyes PRECAUTION Do not cover the eye recognition sensors because covering th...

Страница 8: ...onometer to measure your eye pressure you can remove the probe by powering off the device and allowing the probe to slide out of the probe base Place the probe back in its container and dispose of it...

Страница 9: ...The probe or probe base is dirty or damaged Battery light is flashing Cause Battery power is low ERROR MESSAGE 1 Make sure you are facing straight ahead i e head held at a 90 angle 2 Adjust the tonom...

Страница 10: ...d batteries and insert 2 new CR123A lithium batteries Make sure the batteries positive end is facing the battery cover Rechargeable batteries should not be used Slide the battery cover back over the b...

Страница 11: ...damp cloth or paper towel Make sure the batteries positive end is facing the battery cover PRECAUTION Do not immerse the tonometer in water or any other liquid PRECAUTION Any maintenance services othe...

Страница 12: ...f 5 to 50 mmHg Compared to the Goldmann tonometer less than 5 of measurements fell outside 5 mmHg at each pressure range and less than 1 fell outside 7 5 mmHg at each pressure range The Icare HOME s v...

Страница 13: ...mbined with ISO 13485 certified quality system has made us a respected player in our field of expertise The advanced Icare product line offers reliable high pre cision reproducible accuracy in measuri...

Отзывы: