GUIDANCE AND MANUFACTURER’S DECLARATION–ELECTROMAGNETIC EMISSIONS
Ic
are ic100 (TA011) is intended for use in the electromagnetic environment specified below.
The user of the Icare ic100 (TA011) should assure that it is used in such an environment.
RF emissions CISPR 11
Group 1
Icare ic100 (TA011) is battery operated and
use RF energy only for its internal function.
Therefore, its RF emissions are low and are
not likely to cause any interference in nearby
equipment.
RF emissions CISPR 11
Class B
Icare ic100 (TA011) is suitable for use in
all establishments, including domestic
establishments and those directly connected to
public low-voltage power supply network that
supplies buildings used for domestic purposes
Harmonic emissions
IEC 61000-3-2
NOT APPLICABLE
Voltage fluctuations
flickering emissions
IEC 61000-3-3
NOT APPLICABLE
GUIDANCE AND MANUFACTURER’S DECLARATION–ELECTROMAGNETIC IMMUNITY
Icare ic100 (TA011)
is intended for use in the electromagnetic enviroment specified below.
The customers or users of Icare ic100 (TA011)) should assurre that it is used in such an enviroment
Immunity test
IEC 60601 Test level Compliance level
Electromagnetic environment-
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
±8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %
Electrical fast
Transients/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
NOT
APPLICABLE
NOT APPLICABLE
Surge
IEC 61000-4-5
±1 kV for line(s) to
line(s)
±2 kV for line(s) to
earth
NOT
APPLICABLE
NOT APPLICABLE
Voltage dips, short
interruption and
voltage variations on
power supply lines
IEC 61000-4-11
<5 % UT (>95 %
dip in UT) for
0,5 cycle
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT (30 %
dip in UT) for
25 cycles
<5 % UT (>95 %
dip in UT) for 5 s
NOT
APPLICABLE
NOT APPLICABLE
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.