24
The time required for sterilizer to be ready for routine use after the power is switched is less than 15
minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C
Programs
(STE-23-D)
Tem
per
at
ur
e
(°C
)
Pr
essur
e
(kPa)
Ho
ldi
ng
ti
m
e
(m
in
)
Tot
al
ti
m
e
(m
in)
Type
Ma
x
lo
ad
(
kg
)
Ma
x
lo
ad
p
er
tr
ay
(k
g)
SOLID
134
210
4
25-45
Unwrapped solid material
5.00
1.50
121
110
20
30-60
WRAPPED
134
210
4
35-60
Unwrapped solid material
5.00
1.50
121
110
20
35-65
Single-wrapped solid or
hollow
material
4.00
1.20
TEXTILE
134
210
8
45-65
Unwrapped porous material
1.25
0.40
Single-wrapped porous
material
1.10
0.30
121
110
30
50-75
Dual-wrapped porous material 0.75
0.25
Single-wrapped hollow
material
4.00
1.25
Dual-wrapped solid and
hollow
material
2.00
0.60
PRION
134
210
18
50-75
Unwrapped porous material
1.25
0.40
Single-wrapped porous
material
1.10
0.30
Dual-wrapped porous material 0.75
0.25
Single-wrapped hollow
material
4.00
1.25
Dual-wrapped solid and
hollow
material
2.00
0.60
LIQUID
(optional)
134
210
10
35-60
Liquid
1.20
0.40
121
110
30
35-65
Drying
(optional)
―
―
―
1-20
―
―
―
B&D test
134
210
3.5
28-35
―
―
―
Helix test
134
210
3.5
28-35
―
―
―
Vacuum test
―
―
―
20-25
―
―
―