ICANCLAVE STE-8-D Скачать руководство пользователя страница 14

 

14 

Data 

 
The internal memory will store the information of the last 9999 cycles.  

 

USB Flash memory (Optional) 

 
A USB drive can be used as a method of storing a report of the cycle. To do so, insert the USB drive 
into the slot located on the service door of the sterilizer. 
The information will automatically output directly to the USB drive after the cycle has completed. The 
name of the file is determined by the serial number of the machine and the cycle number. 
For example: 
The serial number is E00001.  The cycle number is 0012. 
The file name in the USB stick is 01001200.txt.   
The first two numbers represent machine number. 
The middle four numbers represent cycle number. 
The last two numbers represent error code. 
E.g. 00:no error; 01: error E01  

 

Printer (Optional) 

If installed, you can see the icon in the screen stop flashing. 
At the end of each cycle the printer will print out a report of the cycle. 
Note: is there is no paper inside the printer; the icon will flash.  
 

Report  

 
Internal Memory 
In this menu you can get the information of all the cycles stored in the internal memory of the sterilizer. 
Select “Report” from the main menu and press   button, you will see the list of records. 
Select the record by pressing   button. 
Press   button to print and save the report. 
Press   button to exit. 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Program 

Basic Set 

Report 

Label  

00012 

00011 

00010 

00009 

Содержание STE-8-D

Страница 1: ...Instructions Manual For Models STE 8 D STE 12 D STE 18 D STE 23 D STE 29 D...

Страница 2: ...wer outage in the manual Need maintenance If this picture appears on the screen when power On or E88 appears on the report please call your dealer or local maintenance service Your steam sterilizer ne...

Страница 3: ...terilization chamber 12 Select the program 12 Start the sterilization program 12 End of cycle 12 Manually interruption of the cycle 13 Test Program 14 Data 14 Save report 14 Printer optional 16 Labels...

Страница 4: ...nnect only to a properly grounded outlet Do not cover or block any openings on this appliance Use this appliance only for its intended use as described in this manual Do not exceed the maximum weight...

Страница 5: ...10 Printer 17 Power socket 3 Door lock 11 USB port 18 Used water tank vent 4 Door handle 12 Main switch 19 Condenser vent 5 Chamber 13 Distilled water outlet 20 Bacteriological filter 6 Door switch Di...

Страница 6: ...it breaker F16A 400 V Operation temperature 5 C 40 C Operation relative humidity Max 80 non condensing Max Noise level 70 dB Atmospheric pressure 76 kPa 106 kPa Packing content Item Accessories Quanti...

Страница 7: ...id containers or other objects on the sterilizer Power supply connection Check the label on the back panel of sterilizer to verify the voltage rating for the unit Failure to connect the sterilizer to...

Страница 8: ...instruments by ultrasound cleaner Residual chemicals left over after cleaning and disinfecting process may damage and corrode parts of the sterilizer always rinse off the instruments using distilled w...

Страница 9: ...et The Setup menu permits to set the following options Parameter Unit Preheat Expire date labels Water quality sensor and see the information of Last error Select Setup from the main menu by pressing...

Страница 10: ...reheat When this mode is activated the chamber and steam generator start to warm until it reaches the minimum temperature to begin a sterilization cycle this helps to reduce the total cycle time and t...

Страница 11: ...process the most relevant information corresponding to the last error can be displayed on the screen Select Last error from the menu by pressing button Press button to back 5 Operation Prerequisites...

Страница 12: ...eted End will appear at the end of the graphic the printer will print out and the digital report saved in the USB memory is these are connected Caution Always use the tray handle to load or unload the...

Страница 13: ...ct Program from the main menu by pressing then to enter in the menu select B D test in the screen will appear the information as the temperature and sterilization time holding time also the date time...

Страница 14: ...200 txt The first two numbers represent machine number The middle four numbers represent cycle number The last two numbers represent error code E g 00 no error 01 error E01 Printer Optional If install...

Страница 15: ...7C 053 7kPa T5 12 43 37 087 9C 075 0kPa T6 12 50 40 134 8C 206 0kPa TS 134 7C 209 5kPa Max Temperature 135 2C Min Temperature 134 3C Max Pressure 214 0kPa Min Pressure 204 9kPa T7 12 54 39 134 4C 211...

Страница 16: ...ect the main cable Drain the tank completely using the drain connector at the back of the sterilizer and leave it connected into the connector in an open position Clean the internal surface with a sof...

Страница 17: ...hten the sealing plate Turn the nut until the sealing plate is tight If the door knob is too tight you may also turn the nut clockwise to loosen it Caution Never adjust the chamber door while the door...

Страница 18: ...t work E11 Failure to preheat the steam generator Power off run a new cycle Contact your supplier if error persists E12 Failure to preheat the chamber Power off run a new cycle Contact your supplier...

Страница 19: ...ase of the steam and restoration of the safety pressure 4 Safety micro switch for the door status Comparison for the correct closing position of the door Action Signal of the wrong position of the doo...

Страница 20: ...icium oxide sio2 1mg I 1 0mg kg Iron 0 2mg I 0 1mg kg Cadmiun 0 005mg I 0 05mg kg Lead 0 05mg I 0 1mg kg Rest of heavy metals 0 1mg I 0 1mg kg Chloride 2mg I 0 1mg kg Phosphates 0 5 mg I 0 1mg kg Cond...

Страница 21: ...erial 2 00 0 60 121 110 20 45 50 Single wrapped solid or hollow material 1 50 0 50 TEXTILE 134 210 8 30 45 Unwrapped porous material 0 50 0 15 Single wrapped porous material 0 35 0 10 121 110 30 55 60...

Страница 22: ...20 45 50 Single wrapped solid or hollow material 2 50 0 80 TEXTILE 134 210 8 30 45 Unwrapped porous material 0 75 0 25 Single wrapped porous material 0 50 0 15 121 110 30 55 60 Dual wrapped porous mat...

Страница 23: ...20 45 60 Single wrapped solid or hollow material 3 50 1 10 TEXTILE 134 210 8 35 50 Unwrapped porous material 1 00 0 30 Single wrapped porous material 0 80 0 25 121 110 30 55 70 Dual wrapped porous mat...

Страница 24: ...20 35 65 Single wrapped solid or hollow material 4 00 1 20 TEXTILE 134 210 8 45 65 Unwrapped porous material 1 25 0 40 Single wrapped porous material 1 10 0 30 121 110 30 50 75 Dual wrapped porous mat...

Страница 25: ...0 WRAPPED 134 210 8 45 60 Unwrapped porous material 3 5 2 0 Single wrapped porous material 3 0 2 0 Dual wrapped porous material 2 5 4 0 2 0 2 0 121 110 30 50 65 Single wrapped hollow material 3 5 2 0...