SPECIFICATIONS
Frequency
Effective radiating area (ERA)
Mode
Work Cycle
Frequency of pulse repetition
Transducer of 7 cm
2
Treatment time
Output Power
Crystal of 7 cm
2
Maximum Amplitude (7 cm
2
)
BNR
Crystal of 7 cm
2
1.1 MHz, ± 10%
3.3 MHz, ± 10%
7 cm
2
Continuous
Pulsed
20%; 50%
100 Hz; 16 Hz; 48 Hz
1.1 MHz ±10%; 3.3 MHz ± 10%
1-30 min
0.1 to 3.0 W/cm²; 1.1 MHz e 3.3 MHz
21 W ± 20%
3
Note: The equipment in pulsed mode shows
values of peak power, average values are
equivalent to pulsed selected percentage, ex:
Selected:
Potency: 21 W
Duty cycle: 50%
Average Power:
21 x 0.5 = 10.5 W