Calibration
49
Measuring of the cell parameter
Before this function can be used, the reference temperature must be
adjusted. It is sufficient for the conductivity calibration to know the cell
parameter. If this is not known, or has been changed after longer use by
deposits on the electrode, this function can be used to define the cell
parameter newly. Call the function
measure cellpar.
in the adjust-menu.
In this function the measuring range will not be adjusted automatically. The
adjustment for dialysate should be in the area of 13...14 mS/cm. Therefore
the measuring range must be 0...20 mS/cm. If the measuring range is too
small, the device indicates the overflow with >
OFL
< above the beam
display. Via the menu point
measure conduct
you can choose the correct
measuring range with the arrow-keys. During the calibration the standard
solution should have a temperature of approx. 25 °C. When the
measurement value in the display is stable, the conductivity value of the
standard solution can be entered via the numeric keyboard and confirmed
with >
ENTER
<. After the calibration, all values relevant for the
conductivity measurement are indicated once more. With the
ok
-key the
data will be saved.
For calibration of the conductivity use only fresh standard
solutions. Used solutions should not be refilled into the storage
bottle! Keep the standard solutions in a cool place.
Содержание HDM99
Страница 1: ...HDM99 Dialysis Measuring System User Manual ...
Страница 15: ...Measurement 15 Measurement From the main menu over measuring you will receive the following menu ...
Страница 21: ...Measurement 21 Conductivity and temperature measurement ...
Страница 26: ...Measurement 26 pH Measurement ...
Страница 46: ...Calibration 46 ...
Страница 68: ...Index 68 V Voltage 6 9 10 36 37 38 39 Y y range 18 19 22 39 ...