8.2 Electrical specifications
Volt
120 VAC, alternating current
Amperage
3.9A
Actuators
24 VDC, direct current
Frequency
60 Hz
Hospital grade plug
Yes
Electrical protection
Class II
Applied part
Type B
Battery type (option)
Rechargeable sealed lead acid
Battery capacity (option)
1.3 Ah
Protection class (IP)
Control box : IPX6
Actuator : IPX4
Lifting column : IPX0
Hand control : IPX6
Foot switch : IPX6
Battery : IPX4
Duty cycle
Intermittent 2/18
Max 2 minutes on
18 minutes off
IBIOM Instruments Ltd. reserves the right to change specifications without notice.
8.3 Environmental conditions
Conditions
Temperature
Relative humidity
Pressure
Normal condition
5
o
C to 40
o
C
20% - 80%
70-106 kPa
Storage
-10
o
C to 50
o
C
20% - 80%
70-106 kPa
Transport
-10
o
C to 50
o
C
20% - 80%
70-106 kPa