9
SETTINGS
User adjustable settings.
Batt. Low Ind.
Battery voltage at which the unit displays BATTERY LOW.
Batt. Low Shutdown
Battery voltage at which the unit will auto-shutdown.
Batt. Chg. Enable
Enable / disable battery charging.
This setting is useful when an external DC source is used but the user
doesn’t want the batteries to charge.
Batt. Chg.
O’L mA
Charge overload current setting.
The charging circuit is self-contained, but as a safety setting to protect the
batteries & electronics if the charge current goes too high (a faulty battery
perhaps). The charging will be shutdown.
The maximum current should be approx. 250mA, so the default setting here
of 400mA gives some margin.
Batt. Chg. Full mA
Current (mA) at which the battery is determined to be near full.
The charging circuit is self-contained, but as an added extra the software can
shutdown charging if the battery is approaching near full capacity. This is
done by monitoring the charge current and if it drops below a specific level
then the charger is shutdown.
B.Light Enable boot
Selects whether the backlighting is enable by default at start up.
The backlighting on/off control on the main menu still allows for the
backlighting to be controlled for that session.
Use the menu options as detailed at the bottom of the screen to navigate the various entries, Quit, Save
etc. Please note that LOAD.DEF will load pre-set factory settings and NOT your as-shipped custom
settings.
See the Calibration record at the back of this manual for all the custom settings for your unit.
Any changes to settings are saved to EEprom.
Note: If you decide to make any adjustments to the above settings then please record the original settings
before you start.