i4Technology 62252 Скачать руководство пользователя страница 5

 

 

Fig. 2

 –

Main parts scheme 

 
1.4.2. The radiation detector used in the device is a gas-discharge meter. By it the photon 

flux is transformed into the sequence of electrical signals by the radiation detector. These signals 
are  processed  by  the  smart  detection  circuit  that  provides  the  automatic  processing  of  the 
ambient dose and the ambient dose rate measured data and the display indication.  

 
1.4.3. Figure 3 illustrates the display in the Measurement mode 
 

 

Fig. 3 – Measurement mode on the display 

 
During  the  measuring  averaged  values  of  the  ambient  dose  rate  is  being  continuously 

displayed  (no.1)  in  the  units  defined  by  user  (µSv/h  or  µR/h).  When  the  radiation  level  is 
constant,  the measuring  time  is ranging  from  10 to 80 seconds depending  on the radiation rate. 
The displayed values are automatically changing and being averaged by the microprocessor. This 
averaging process is displayed by the measurement cycle dynamic graph (no. 5). 

 
The additional information shown on the display includes also: 

 

radiation environment time base diagram (no. 2); 

 

environment temperature (по no. 6); 

 

sound mode pictograph (no..7); 

 

pictograph informing about the chosen power type (no. 8); 

 

battery charge level (no. 9); 

 

current time (no. 10); 

 

textual  information  about  the  exceedance/nonexceedance  of  the  ambient  dose  rate 
threshold defined by user (no.11). 
 

Note.  The  environment  temperature  is  measured  by  the  inbuilt  sensor.  For  this  reason 

displayed value of  the environment  temperature  may depend on the heating of  the unit’s parts 
while operating a device.  

 
In the line (no. 3) below the radiation environment time base diagram (no. 2) you can see 

Содержание 62252

Страница 1: ...Package and transportation 6 Proof of acceptance 7 Manufacturer s warranty 8 Certification information Warranty Card Appendix General information of the radiation nature and the material radiation in...

Страница 2: ...ifferent readings This situation is similar to the rain in everyday life When there is a bit of rain it s difficult to detect how many drops drip per second by hand there may be 2 drops more or no one...

Страница 3: ...accumulated dose selecting the unit measures roentgen Sievert setting and changing the alarm sound of the ambient dose rate threshold audiovisual warning of the ambient dose rate threshold excess sett...

Страница 4: ...re than 105 58 18 5 Mass with batteries grams not more than 110 1 3 Contents of the box 1 3 1 The table below gives the information about parts of the product and the delivery set Table 2 Name number...

Страница 5: ...nging from 10 to 80 seconds depending on the radiation rate The displayed values are automatically changing and being averaged by the microprocessor This averaging process is displayed by the measurem...

Страница 6: ...iation dose considered safe During the life the accumulated dose value must be below 100 700 millisieverts summarized according to the local radiation background 1 4 4 Control buttons are used to swit...

Страница 7: ...edure To prepare the product for working you should loosen the screw and remove the battery compartment cover no 3 Fig 2 install the batteries respecting polarities Fig 4 place back the battery compar...

Страница 8: ...button come back to the Measurement settings menu Fig 8 for further settings 2 3 1 4 Setting the threshold of the ambient dose rate above which the device sounds because the radiation level is high wh...

Страница 9: ...opposite the chosen brightness and the screen brightness will be changed according to your choice by pushing the button come back to the Screen settings menu fig 9 for further settings 2 3 1 8 Setting...

Страница 10: ...hing the buttons choose the line BUTTONS SOUND and confirm your choice by pushing the button or by pushing the buttons and choose the value ON or OFF to switch the buttons sound on and off and confirm...

Страница 11: ...rgy of the batteries when user forgets to switch off the device For settings you should when you are in the Main menu 7 by pushing the buttons choose the line SETTTINGS and confirm your choice by push...

Страница 12: ...e the line SETTTINGS and confirm your choice by pushing the button or in the opened submenu by pushing the buttons and choose the line ACCUMULATOR and confirm your choice by pushing the button or by p...

Страница 13: ...dio signal passing frequency occur At the same time the ambient dose value is raising in the upper right corner of the screen and the columns at the radiation environment time based historical graphs...

Страница 14: ...e the natural radiation background normal for the location it indicates that the object you investigate is polluted by radiation To define the ionization radiation source you should remove the working...

Страница 15: ...pair should be carried out by the dealer 5 Package and transportation Every Product delivery set see Table 2 is packaged in individual corrugated fiberboard box The moving of contents of the box is no...

Страница 16: ...or the authorized Distributor shall replace or repair the products broken down during the guaranteed use period at the expense of them 7 5 This warranty shall be void if guaranteed use period is over...

Страница 17: ..._________________________________________________________________________ Sale Date _________________________________________________________________________ Vendor s stamp and signature _____________...

Страница 18: ...es caused to biological tissues by ionizing radiation exposure depend on the dose of absorbed radiation energy To characterize it the term absorbed dose is used i e the energy absorbed by the unit of...

Страница 19: ...The equivalent dose rate H is defined by the formula H dH dT where dT is a period of time In the SI system the equivalent dose rate H is measured by Sievert per second Sv s 1 Sv s 100 Rem s 1 Sv s R s...

Отзывы: