WARNING:
→
Keep at a proper distance the electrodes during treatment.
Electrodes that come in contact with each other could lead to
unwanted stimulations and/or burns.
→
The current intensity depends on the electrodes’ dimensions.
Improper application can lead to damage to the patient. In case
of any doubts on the proper electrodes’ size, consult a
physician/therapist specialized in electrostimulation.
This device is supplied with 50×50mm and 50×100mm adhesive electrodes.
You can select the correct adhesive electrodes according to the treatment area
and output current density.
It is recommended to use only electrodes supplied by the manufacturer, to
guarantee the highest level of contact with the area to be treated and the right
current density in relation with the prescribed treatment. Properly dispose of
electrodes after the use.
In case of doubts about the electrodes’ integrity, it is suggested to replace them
with new ones.
Do not start the treatment before the electrodes are correctly placed on the
skin. Do not take away the electrodes during the ongoing therapy.
Pre-gelled electrodes positioning
1.
Insert the cable with the Red (+) connector into
one adhesive electrode. Insert the cable with the
Black (-) connector into one adhesive electrode.
Make sure that the connectors are inserted
completely and there are no metal parts of the
pins exposed.
2.
Remove the adhesive electrodes from the
protective backing and place on the treatment
area. Make sure that the entire electrode surface
is in contact with patient skin.
As an alternative you can use silicon electrodes for
electrotherapy treatments, in two different ways:
• with the use of wet sponges;
• with the silicon electrodes and conductive gel.