IACER SRL Page 15 di 24 MNPG102-02 30/07/15
These two cycles are repeated in sequence till the end of duration of
the treatment.
5. PRESSURE TIME: Select air flow speed by pressing SET
buttons. You can choose 9 speed levels : 5,7,9,11,13,15,17,19,21
seconds.
Default speed: 7 seconds.
6. INTERVAL: Select interval time by pressing SET
buttons. You
can choose 6 time intervals: 5,10,15,20,25,30 seconds. Default interval
time 0 seconds.
7. TIME: Select time therapy by pressing SET
buttons. Therapy
time can be adjusted from 5 to 99 minutes. Default therapy time 15
minutes.
We suggest to start with treatment of 10-15 minutes and increase it
progressively according to your needs.
Attention: It’s not possible to change therapy time during the treatment.
8. PRESSURE: Select the pressure by pressing SET
buttons. It is
possible to adjust pressure for each compression chamber from 0 to
240 mmHg.
Attention: Start the treatment with a middle pressure and increase it
progressively according to your needs.
9. Press START/STOP button to start the therapy. Any moment it’s
possible to stop the treatment pressing again START/STOP button.
10.Press RESET
button to delete adjusted parameters and restore
the default settings.