IACER SRL Page 14 di 24 MNPG102-02 30/07/15
Working instructions
1. Connect cable to the domestic plug 220-240 VAC, then turn ON (1)
the ON/OFF switch (1);
2. Select the parameters by pressing on MENU’ button.
3. Press on MOVE
button to scroll the menu
(MODE, PRESSURE TIME,
INTERVAL, TIME (min), PRESSURE (mmHg)).
4.
MODE. Select the program by pressing on SET
button:
MODE A
1
: Air flows in the first chamber (lower limb) according to
adjusted pressure and sequentially in the other chambers in a distal-
proximal way.
MODE A
2
: Air flows in the fourth chamber (upper limb) according to
adjusted pressure and sequentially in the other chambers in a proximal-
distal way.
MODE B: Air flows in the first two chambers. When adjusted pressure is
reached air flows in the last two chambers (distal-proximal way).
MODE C: Air flows in the first chamber (lower limb) according to
adjusted pressure and sequentially in the other chambers (distal-
proximal way ) by keeping swelled up the previous chamber every time.
MODE D (1-2-3-4) : Air flows only in the selected compression chamber
(mode D1 chamber 1 ; mode D2 chamber 2 ; mode D3 chamber 4 ;
mode D4 chamber 4).
MODE E (MODE A
1
+ MODE C) : The machine runs two cycles in
sequence. During the first cycle air flows in the first chamber (lower
limb) according to adjusted pressure and sequentially in the other
chambers in a distal-proximal way. At the end of the first cycle the
machine starts the second cycle and air flows in the first chamber
(lower limb) according to adjusted pressure and sequentially in the
other chambers (distal-proximal way ) by keeping swelled up the
previous chamber every time.