background image

 

I.A.C.E.R. Srl 

                    MNPG207-02 

       

 

Applied part type BF 

 

Compliance with Directive 93/42/EEC and following 
modifications 

 

Manufacturing date (month/year) 

 

Admission temperatures 

 

 

Relative humidity 

 

Fabricant  information 

IP22 

Degree of protection against ingress of solids , powders and 
liquids (device protected against solid foreign diameter 

12,5mm and against dripping water vertical when the device 

is kept at 15 ° from the normal operating position ) . 

 

Only domestic use 

Kit contents 

ORTHOMAG kit: 

 

N°1 ORTHOMAG device; 

 

N°1 wall mount charger (cable 1.5 m); 

 

N°1 user and maintenance manual; 

 

N°1 universal flexible applicator (cable 1.5 m); 

 

N°1 carriage bag; 

 

N°1 test emissions magnet; 

 

N°2 elastic bands (S and L size); 

 

N°1 car charger (optional). 

 
Visit website 

www.orthomag.eu

  to obtain more information. 

 

Содержание ORTHOMAG

Страница 1: ...USER MANUAL MNPG207 02 Edition 02 12 2016 Magnetotherapy model ORTHOMAG I A C E R Srl www itechmedicaldivision com...

Страница 2: ...I A C E R Srl Via S Pertini 24 A 30030 Martellago VE ITALY Tel 39 041 5401356 Fax 39 041 5402684 e mail iacer iacer it www itechmedicaldivision com...

Страница 3: ...r 5 Declaration of conformity 5 Specifications 6 Technical features 7 Labelling 8 Label details 8 Symbols description 8 Kit contents 9 How to use 10 Warnings 10 Contraindications and side effects 12 H...

Страница 4: ...rticular for the osteoporosis Pulsed electromagnetic fields induce biological modifications on biological membrane in order to re establish correct cellular functions According to different authors ex...

Страница 5: ...bility that ORTHOMAG is manufactured in conformity with Directive 93 42 EEC MDD dated 14 June 1993 D Lgs 46 97 dated 24 February 1997 Attuazione della Direttiva 93 42 CEE concernente i dispositivi med...

Страница 6: ...or when mixed with any flammable agents and in environments with high concentrations of oxygen Continuous operating mode equipment Equipment not suited to be used in external Purpose Clinical purpose...

Страница 7: ...e of the device and its accessories 3 years Technicalfeatures Power supply Battery charger DC 3 7V Lithium battery 12V 1A Max absorbed current 300 mA therapy mode Insulation class CEI EN 60601 1 II Ap...

Страница 8: ...etails Label 1 Label 2 DC INPUT 12V 1A Symbolsdescription Attention consult operating instructions Product subject to WEEE regulations concerning separate waste collection of electronic equipment Clas...

Страница 9: ...e protected against solid foreign diameter 12 5mm and against dripping water vertical when the device is kept at 15 from the normal operating position Only domestic use Kitcontents ORTHOMAG kit N 1 OR...

Страница 10: ...o not open the device in case of problems contact the manufacturer Orthomag is a medical device Keep it out of children and animal s reach Do not use the device in case of anomalies or malfunctions Me...

Страница 11: ...ay from other devices that may interfere or prevent the safe use of the device and its accessories connected The manufacturer is considered responsible for the performance reliability and safety of th...

Страница 12: ...tes viral in acute phase illnesses mycosis cardiopathic subjects tumours serious arrhythmias or pacemaker carriers children metallic prosthesis carriers acute infections epileptics different medical p...

Страница 13: ...device switches off When using the device with battery the program duration is guaranteed up to 3 hours How to use the device with the power supply adapter Connect the applicator to the Orthomag Place...

Страница 14: ...field is still emitted until the device is switched off Therapeutic indications average time of therapy 2 4 hours per day To discontinue therapy and switch off the device press and hold the power but...

Страница 15: ...tor allows a perfect contact with the body The image below gives some of the possible applications for the most common diseases treated with magnetic therapy such as cervical osteoarthritis and or joi...

Страница 16: ...I A C E R Srl 16 MNPG207 02...

Страница 17: ...device from the charger Remove the belt clip by pushing it down lightly Open the battery compartment Pull out the battery Insert the new battery please use only battery provided by the manufacturer Cl...

Страница 18: ...carefully using a sponge soaked in solution of water and alcohol 20 We recommend to disconnect the applicator from the device before cleaning the it Pay attention to respect the temperature humidity a...

Страница 19: ...AG please follow these instructions Check the mains socket integrity by connecting a working device to the same socket Check the connection with working wall mount charger and applicator Check the cor...

Страница 20: ...0030 Martellago VE Tel 041 5401356 Fax 041 5402684 You can get any technical documentation on spare parts but only prior business authorization Spareparts For original spare parts contact the National...

Страница 21: ...er supply network that supplies buildings used for domestic purposes Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Electromagnetic immunity ORTHOMAG is intended for use in the electrom...

Страница 22: ...r of the ORTHOMAG requires continued operation during power mains interruptions it is recommended that ORTHOMAG be powered from an uninterruptible power supply or a battery Mains power electromagnetic...

Страница 23: ...han the complicance level in each frequency rage Interference may occur in the vicinity of equipment marked with the following symbol Recommended separation distances between portable and mobile commu...

Страница 24: ...g to the transmitter manufacturer Note 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies 2 These guidelines may not apply in all situations Electromagnetic propaga...

Страница 25: ......

Страница 26: ......

Страница 27: ......

Страница 28: ......

Отзывы: