background image

 

 

 

 

Содержание Mio-Sonic

Страница 1: ...USER MANUAL Ultrasound therapy MIO SONIC...

Страница 2: ......

Страница 3: ...ESCRIPTION 5 LABELLING 5 Package content 6 HOW TO USE 7 INTRODUCTION TO THE TECHNOLOGY 7 CONTRAINDICATIONS 9 Side effects 9 WARNINGS 9 DEVICE USE 11 Antalgic and de contracting treatments 14 Beauty tr...

Страница 4: ...Directive 93 4 EEC transposed in Italy by the D Lgs 46 97 as modified by the Directive 2007 47 EC D Lgs 37 2010 and further modifications integrations The product has been assigned to class IIa accord...

Страница 5: ...s ideal for the treatment of muscular and nervous pathologies and for the recovery of traumas in case of both chronic and acute pathologies In fact ultrasound therapy is indicated for antalgic treatme...

Страница 6: ...lation 100Hz 10 Waveform Pulsed Modulation duty cycle L 5 M 50 H 100 Regolation intensity Adjustable on 3 steps L M H Ultrasound head 5 cm Effective radiant area 4 cm 20 Ultrasonic beam Collimated Hea...

Страница 7: ...NPG71 06 Device and commands description 1 Time indicator light 2 Time button 3 Power indicator light 4 Intensity indicator light 5 Mode button 6 Power switch 7 Ultrasound head 8 AC DC adapter connect...

Страница 8: ...for the disposal of electronic waste Applied part type BF Class II device IPX7 Degree of protection from temporary immersion under water up to 1m of depth and up to 30 minutes only for the ultrasound...

Страница 9: ...gy angiology and cardiology these technologies in fact exploit the return echo deriving from an ultrasonic beam which propagates inside the human body and which is slowed down in a different way depen...

Страница 10: ...ts into formation of steam zones within the fluid The dissolved gasses within the fluid they aggregate as a result of the lowering of pressure realized by the ultrasound forming bubbles or cavity cont...

Страница 11: ...s from epilepsy phlebitis in place thrombophlebitis in feverish state anxiety or serious illnesses vein thrombosis severe osteoporosis inflammation arteriopathies except in case of medical prescriptio...

Страница 12: ...n environments with high concentrations of oxygen with aerosol therapy devices or in wet environments use of the device is prohibited in bathroom or shower areas or while showering bathing to use the...

Страница 13: ...t immediately and consult the retailer or manufacturer Consult a doctor before using MIO SONIC with metallic osteosynthesis devices IF YOU HAVE ANY DOUBTS REGARDING THE DEVICE USE CONSULT YOUR DOCTOR...

Страница 14: ...reated so that it is free of oil and dirt Apply a generous layer of ultrasound transmission gel on the treatment area The gel is fundamental to ensure a correct fit between the treated area and the ul...

Страница 15: ...M for H intensity is recommended maximum attention in maintaining in continuous moving the ultrasound head The intensity L corresponds approx to 0 08W cm the intensity M approx to 0 80W cm and intens...

Страница 16: ...ycle of therapy with an average duration of 10 minutes each for a maximum period of 21 days In case the pain persists stop for 7 days the treatment and eventually restart another therapy cycle of 21 d...

Страница 17: ...ion of liquids and it is present in the adipose tissue especially arround the ankles the calves and it affects subjects with a good health and with a tonic musculature Flaccid is manifested in middle...

Страница 18: ...will vary from 20 to 30 minutes This job can be performed at intervals of hours or days between an area and the adjacent one or in rapid sequence The use of lotions or gel containing active principles...

Страница 19: ...nee gluteus arms abdomen with maximum ultrasound intensity at M Programs Area Intensity N app tot Frequency Thighs drainage M H 20 Daily Thighs compact cellulite M H 30 Daily Thighs flaccid cellulite...

Страница 20: ...flaccid cellulite L M 20 Daily Arms edematous cellulite L M 20 Daily Abdomen drainage L M 20 Daily Abdomen compact cellulite L M 25 Daily Abdomen flaccid cellulite L M 30 Daily Abdomen edematous cell...

Страница 21: ...e removed using a sponge soaked in solution of water and alcohol Device not subject to sterilization Note Never use solvents for cleaning Cleaning agents cause damage to the device Attention to the ne...

Страница 22: ...erly Check the socket operation Network cable not correctly inserted in the registered connector Insert the plug and cable correctly into the device connector Network cable worn out and interrupted Re...

Страница 23: ...for recycling of electrical and electronic equipment For further information on the obsolete equipment disposal please contact the dedicated disposal service or the shop in which the device was bough...

Страница 24: ...paration or free substitution hand work included 5 Warranty is not applied to damages caused by negligence or use not compliant to the given instructions by intervention on the device from personnel n...

Страница 25: ...es and health establishments According to operating principles the device does not generate significant radio frequency energy and is adequately immune to radiated electromagnetic fields under such co...

Страница 26: ...o 1 Group 1 Il MIO SONIC utilizza energia RF solo per il suo funzionamento interno Perci le sue emissioni RF sono molto basse e verosimilmente non causano nessuna interferenza negli apparecchi elettro...

Страница 27: ...should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Transitori treni elettrici veloci Electrical fast transient burst IEC...

Страница 28: ...5 cicli cycles 5 UT 95 buco in dip in UT per for 5s 40 UT 60 buco in dip in UT per for 5 cicli cycles 70 UT 30 buco in dip in UT per for 25 cicli cycles 5 UT 95 buco in dip in UT per for 5s anche dura...

Страница 29: ...espected calculated from the equation applicable to the frequency of the transmitter Distanza di separazione raccomandata Recommended separation distance RF condotta Conducted RF IEC 61000 4 6 3Veff d...

Страница 30: ...essere previste teoricamente e con precisione Per valutare un ambiente elettromagnetico causato da trasmettitori RF fissi si dovrebbe considerare un indagine elettromagnetica del sito Se l intensit di...

Страница 31: ...aximum output power of transmitter W Distanza di separazione alla frequenza del trasmettitore m Separation distance according to the frequency of the transmitter m 1 2 150 80 150 800 0 35 80 800 80 80...

Страница 32: ......

Отзывы: