Guidance and manufacturer’s declaration – ELECTROMAGNETIC IMMUNITY –
FOR ALL EQUIPMENT AND SYSTEMS
MIO-IONOTENS is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such
an environment.
Immunity test
Test level
IEC 60601
Compliance level
Electromagnetic
environment - guide
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8kV in contact
±2Kv, ±4kV,
±8kV; +15kV on
air
±8kV in contact
±2Kv, ±4kV,
±8kV; +15kV on
air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV
for power
supplies lines
±2kV
for power
supplies lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
±1kV
for input/output
lines
±1kV
for input/output
lines
Impluses
IEC 61000-4-5
±0.5kV, ±1kV
Line(s) to line(s)
±0.5kV, ±1kV
Line(s) to line(s)
Mains power quality
should be that of a
typical commercial or
hospital environment.
±0.5kV, ±1kV,
±2kV
Line(s) to earth
±0.5kV, ±1kV,
±2kV
Line(s) to earth
Voltage dips, short
interruptions and
voltage variations
on power supplì
input lines
IEC 61000-4-11
0% U
T
a 0°, 45°,
90°, 135°, 180°,
225°, 270° and
315° for 0,5
cycles
0% U
T
for 1 cycle and
70% U
T
for 25/30 cycles
at singular phase
0°
% U
T
a 0°, 45°,
90°, 135°, 180°,
225°, 270° and
315° for 0,5
cycles
0% U
T
for 1 cycle and
70% U
T
for 25/30 cycles
at singular phase
0°
Main power quality
should be that of a
typical commercial or
hospital environment.
If the user of MIO-
IONOTENS requires
continued operation
during power mains
interruptions, it is
recommended tha
MIO-IONOTENS be
powered from an
uninterruptible
Содержание MIO-IONOTENS