I.A.C.E.R. Srl
27
MNPG52-04
EMC tables
Electromagnetic emission
Emission test
Compliance
Electromagnetic environment –
guidance
RF emissions
Cispr 11
Group 1
MAG 2000 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment
RF emissions
Cispr 11
Classe B
MAG 2000 is suitable for use in all
establishments,
including
domestic
establishments
and
those
directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Complies
The MAG 2000 is suitable for use in all
establishments, other than domestic
establishments
and
those
directly
connected to the public low voltage
power supply network that supplies
buildings used for domestic purposes
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Electromagnetic immunity
MAG 2000 is intended for use in the electromagnetic environment specified below. The
customer or the user of MAG 2000 should assure that is used in such environment.
Immunity test
Test
level
EN
60601-1-2
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic
discharge (ESD)
EN 61000-4-2
±
6kV contact
±
8kV air
±
6kV contact
±
8kV air
Floors should be
wood, concrete or
ceramic tile. If floor
are covered with
synthetic
material,
the relative humidity
should be at least
30%
Electrical
fast
transient/burst
IEC 61000-4-4
± 2kV for power
supply lines
± 2kV per power
supply lines
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Содержание MAG2000