IACER Srl
28
MNPG179
General indications for a proper use
-
After use, the operator must always turn off and unplug the equipment
and check the condition of the optical fiber and the lens of the laser
output on the handpiece-applicator. In the event of any damage or
impurities, it is recommended to set the device aside and contact the
manufacturer.
-
Handle the handpiece-applicator with care: rough handling can affect
its performance and characteristics. Carry the protected device in the
case provided. The handpiece is connected to the device by means of a
fiber optic: NEVER bend this sharply and NEVER remove or disconnect
it during normal operation, and NEVER tamper with the connection
cable of the handpiece. Failure to observe these requirements can
result in damage to the fibers or to the beam's optical transmission
system and cause injury to the patient or user.
-
The technical staff are not authorized, for any reason, to open and/or
disassemble the handpiece/applicator: this could result in damage to
the handpiece and render the warranty null and void.
-
Do not, for any reason, disassemble the equipment for the purpose of
cleaning or inspection: there is no need to clean inside the machine and,
in any case, the machine should be opened only by the authorized,
specially trained technical staff from IACER Srl.
-
NEVER allow liquids to enter the cavities. NEVER work in environments
that are rich in oxygen. NEVER cover the air vents of the machine,
NEVER immerse the machine in water.
-
Do not spray or pour liquids on the external case of the equipment, in
the air vents, on the LCD display, or in the grate of the fan. If this does
occur, make sure the machine is serviced. IACER Srl cannot be held
responsible for damage due to use of the machine contrary to the
requirements above.
-
If the laser point is missing or of reduced intensity, turn off the device
and contact the technical assistance service.
-
Only authorized technical staff from the manufacturer may access the
internal parts of the equipment. Please contact IACER Srl or its
authorized service centers for repairs and further information.