USER MANUAL

 

MNPG179 Rev. 2 of 03/07/2019 

 

Laser therapy 

 

 

 

 

 
 

 

Содержание I-TECH LA8000

Страница 1: ...USER MANUAL MNPG179Rev 2of03 07 2019 Laser therapy...

Страница 2: ......

Страница 3: ...TECHNOLOGY 11 INDICATIONS 13 CONTRAINDICATIONS 13 WARNINGS 14 PATIENT PREPARATION 16 INSTRUCTIONS FOR USE 16 Warning 16 Connections 18 Start up and protection password 18 Main menu 19 Program selectio...

Страница 4: ...IACER Srl ii MNPG179...

Страница 5: ...ept at hand for quick reference and correctly stored The manufacturer declines all responsibility for improper use of the device use contrary to specific national laws incorrect installation defective...

Страница 6: ...D24021 from the Notified Body no 0476 Kiwa Cermet Italia Spa Declaration of conformity I A C E R S r l Via S Pertini 24 A 30030 Martellago Ve Italia herewith declares under its own responsibility that...

Страница 7: ...on EN 60825 1 equipment unsuitable for use in presence of a flammable anesthetic mixture containing air oxygen and nitrous oxide equipment suitable for continuous operation equipment unsuitable for ou...

Страница 8: ...tt Two delayed type fuses T T2 5A Backlit LCD display 128x64 pixel graphic display Power Settable from 1W to 8W 10W 3 Emission wavelength of laser diode 974nm 3nm Laser classification IEC 60825 1 Clas...

Страница 9: ...he skin Around 38 mm Spot area Around 10mm2 3 Operating conditions environmental temperature From 10 C to 40 C relative humidity From 15 to 93 without condensation atmospheric pressure 500 1060 hPa Co...

Страница 10: ...IACER Srl 6 MNPG179 Device and command description...

Страница 11: ...IACER Srl 7 MNPG179 Labels...

Страница 12: ...cs connector LABEL 4 Placed on one side of the device in warns on the danger of the laser radiation LABEL 5 Placed on one side of the device LABEL 6 Placed on the back of the device it resumes the pri...

Страница 13: ...12 It indicates the emergency stop button LABEL 13 It indicates the position of the laser aperture Symbol Meaning Manufacturer s logo CE product mark released by the Notification Body no 0476 Class of...

Страница 14: ...o wear protective glasses Packaging content The I TECH LA8000 LA10000 box contains n 1 user manual n 1 power supply cable shuko plug n 2 spare fuses see table Technical features n 1 handpiece n 1 peda...

Страница 15: ...therefore permits immediate alleviation of the symptoms of inflammation and degeneration in the fields of orthopedics neurology and dermatology and reduces healing time it is also a valid form of ther...

Страница 16: ...e deep stimulation of tissue for rapid and uniform cell regeneration the I TECH LA8000 10000 permits deep stimulation for treating internal tissue and structures like the articulation of the femur and...

Страница 17: ...t exposure to eyes class 4 lasers are potentially harmful to the retina although the likelihood of damage to the retina is extremely low The operator must however wear the special safety glasses provi...

Страница 18: ...and drugs triggers this response It is recommended that patients at allergic risk or patients with a history of such reactions are tested with a minimum time of treatment Means of fixation metal plate...

Страница 19: ...the I TECH LA8000 After carefully inserting the fiber optic in the socket turn the locking ring clockwise to the end of its stroke without forcing it In any case turn the locking ring to bring the ref...

Страница 20: ...follows 1 clean the skin with care where the head of the laser is to be positioned using either water and soap or alcohol 2 dry the skin thoroughly Cleaning and disinfection must be carried out syste...

Страница 21: ...the energy settings configured on the display For this reason keep the handpiece in contact with the affected area during emission of the laser After activating the handpiece by means of contact with...

Страница 22: ...e socket on the front panel fastening carefully the connector so that the mark on the connector of the handpiece will be aligned with the mark on the front panel of the device ATTENTION do not strain...

Страница 23: ...from 0 4 to 10 mm2 The equipment works most efficiently with the spacer fully extended for a laser output of 10mm2 This special HANDPIECE therefore ensures a broader range of action and a higher degre...

Страница 24: ...igger point mode while the second is in scanning mode The operator is free to choose whether to proceed with the second phase or not the settings of the second phase appear at the end of the first one...

Страница 25: ...returns to the main menu PROGRAMS LIST N Name Watt s Cycle Hz Duty Cycle Time r min Punti p Area cm2 Joule cm2 1 Analgesic 1 1 100 50 4 4p 120J 2 Analgesic 2 5 200 60 5 50cm2 18J cm2 3 Neck pain 1 1 1...

Страница 26: ...cm2 25 Muscle injuries 4 500 30 5 25cm2 14J cm2 26 Meniscopathy 2 500 50 2 5cm2 24J cm2 27 Sinovitis 4 1000 30 6 25cm2 17J cm2 28 Muscle Strain 6 500 30 10 25cm2 43J cm2 29 Tendinopathy 2 100 40 2 5cm...

Страница 27: ...f the LA10000 This parameter does not change in relation to the other parameters Cycles For enabling and modulating the laser emission The first choice is Continuous where the emission of the laser is...

Страница 28: ...mit the laser it permits a pause with the actuator not pressed of no more than 3 minutes After this length of time the device disables the laser and returns to the main menu ENTER key during therapy t...

Страница 29: ...ime in order to determine the diode s lifetime It is a parameter useful to the manufacturer used for the device electronic problems resolution Device scare Maintenance For safety reasons ALWAYS turn o...

Страница 30: ...ore carrying out any maintenance or cleaning operations It is strictly recommended not to use diluents detergents acid solutions aggressive solutions or flammable liquids to clean the external surface...

Страница 31: ...to replace correctly the fuses on the back of the device follow the instructions below turn off the equipment unplug the equipment use a screwdriver to open the fuse holder tray making sure to insert...

Страница 32: ...mble the handpiece applicator this could result in damage to the handpiece and render the warranty null and void Do not for any reason disassemble the equipment for the purpose of cleaning or inspecti...

Страница 33: ...nnector of the equipment The power cable is worn and damaged Replace the power cable The emergency switch is turned off Turn on the emergency switch One or more of the fuses are defective or damaged R...

Страница 34: ...s of pressure in the actuator Check air tightness When any of the following conditions occur unplug the equipment and contact the IACER Srl technical assistance service the cable or rear integrated po...

Страница 35: ...or further information on the obsolete equipment disposal please contact the dedicated disposal service or the shop in which the device was bought Warranty IACER Srl guarantees a warranty period unles...

Страница 36: ...which have been in contact with patients Any equipment which the technical department does not consider hygienic Italian law D Lgs 81 2008 on safety in the workplace will not be accepted 3 disassembl...

Страница 37: ...nstalled in homes and health establishments The equipment does not generate significant radio frequency energy and is adequately immune to radiated electromagnetic fields Therefore it does not detrime...

Страница 38: ...ns CISPR 11 Group 1 The device must emit electromagnetic energy at RF in order to perform its intended function So its RF emissions are very low and therefore it does not affect electronic equipment p...

Страница 39: ...material the relative humidity should be at least 30 Transistor Electri cal fast transient IEC 61000 4 4 2kV for power supply lines 2kV per power supply lines Mains power quality should be that of a t...

Страница 40: ...environment guidance 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT per 5s 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT per 5s power mains interruptions it is recommended that the device be...

Страница 41: ...ation distance Conducted RF IEC 61000 4 6 3Vrns from 150kHz to 80MHz 6Vrms from 150kHz to 80MHz in ISM band 3Vrns from 150kHz to 80MHz 6Vrms from 150kHz to 80MHz in ISM band P V d 1 5 3 P V d 1 12 for...

Страница 42: ...such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess...

Страница 43: ...800 ISM band 80 800 800 6 to RF wireless radio communication equipment 0 01 0 12 0 2 0 12 0 23 0 1 0 38 0 63 0 38 0 73 0 2 1 1 20 2 0 1 20 2 30 1 8 2 10 3 80 6 3 3 80 7 30 100 12 00 20 12 00 23 00 Fo...