IACER SRL Page 22 MNPG219-00 25/03/16
Electrical
fast
transient/burst
IEC 61000-4-4
± 2kV for power
supply lines
± 2kV per power
supply lines
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Impulses
EN 61000-4-5
±
1kV
differential
mode
±
1kV
differential
mode
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Voltage dips, short
interruptions
and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 5% U
T
(>95% dips of U
T
)
per 0,5 cycles
40% U
T
(60% dips of U
T
)
per 5 cycles
70% U
T
(30% dips of U
T
)
per 25 cycles
< 5% U
T
(>95% dips of U
T
)
per 5 seconds
< 5% U
T
(>95% dips of U
T
)
per 0,5 cycles
40% U
T
(60% dips of U
T
)
per 5 cycles
70% U
T
(30% dips of U
T
)
per 25 cycles
< 5% U
T
(>95% dips of U
T
)
per 5 seconds
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
If the user of the I-
PRESS 4 requires
continued operation
during power mains
interruptions, it is
recommended that
I-PRESS 4 be
powered from an
uninterruptible
power supply or a
battery.
Mains
power
electromagnetic field
EN 61000-4-8
3 A/m
3 A/m
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.