background image

40

   www.i-sens.com

www.i-sens.com  

41

⑦  

Repeat the same procedure for the 

lancing device (step 

 to step 

).   

*  After the pre-cleaning and disinfection procedure, 

the control solution should be tested to confirm that 

the meter works properly before using the meter. 

Control solution tests should be performed with two 

different levels of control solutions (Level A and B). 

Verify that the test results are within the range printed 

on the test strip box. See pages 22–23 for how to do a 

control solution test.

Note: 

If any of the following deterioration signs appear

after pre-cleaning or disinfecting, please stop using the

system and contact Customer Service at 1-800-429-

5001 Mon–Sat, 9 am–9 pm EST. 

At all other times, 

contact your healthcare professional. 

•   When the inscriptions on the exterior of the meter (or 

lancing device) have been removed

•   When the color of the meter (or lancing device) has 

changed or faded

•   When cracks or roughness develop on the meter (or 

lancing device)

•   When a part of the segment on the meter display does 

not flash

•   When control solution test results do not fall within the 

stated range on the test strip box.

Caution: 

•   Do not use other cleaners or disinfectants because 

other chemicals have not been validated and may 

damage the meter.

•   Do not get fluids inside the meter through the test strip 

port, data transmission port or battery compartment. 

Never immerse the meter or hold it under running 

water because this will damage the meter.

Caution: Storage and Handling

•   Do not expose the meter to direct sunlight, heat, or 

excessive humidity for an extended period of time.

•   Do not let dirt, dust, blood, or water enter into the 

meter’s test strip port.

•   Do not drop the meter or subject it to strong shock.

•   Do not try to fix or alter the meter in any way.

•  Keep the meter in a cool and airy place. 

•   Store all meter components in the carrying case to 

prevent loss.

Note: 

You can get additional information or technical 

assistance by calling our Customer Service Center: 

1-800-429- 5001 (9 am–9 pm EST, Monday–Saturday). 

At 

all other times, contact your healthcare professional. 

Содержание KetoSens

Страница 1: ...tting the Strip Expiration Date Indicator 20 Checking the System 21 Control Solution Testing 22 Comparing the Control Solution Test Results 24 Testing Using the Lancing Device 25 Preparing the Lancing...

Страница 2: ...performing correctly Please use this device only for the intended use described in this User s Manual Please follow the suggested cleaning and disinfection procedures described in this User s Manual...

Страница 3: ...humidity may affect results For single patient use only For over the counter use Important Information The KetoSens Blood Ketone Monitoring System is intended for self testing outside the body in vitr...

Страница 4: ...and B are sold separately and can be purchased by contacting Customer Service Check all the components after opening the KetoSens blood Ketone monitoring system package The exact contents are listed...

Страница 5: ...r as shown in the figure on the right Insert two new batteries with the side facing up and make sure the batteries are inserted firmly Step 3 Place the cover on the battery compartment Push down until...

Страница 6: ...strips in any way For detailed storage and usage information refer to the KetoSens test strip package insert Keep the meter and testing supplies away from young children The test strips and foil packe...

Страница 7: ...to turn off the meter Press the or button to change values Press and hold the or button to scroll faster Adjusting the Date and Time Step 1 Entering SET Mode Press and hold the S button for 3 seconds...

Страница 8: ...he date and to go to the next step Step 5 Setting the Time Format The meter can be set in the AM PM 12 hour or the 24 hour clock format Press the or button to select a format The AM PM symbol is not d...

Страница 9: ...etting the Sound On OFF Note The symbol is displayed only when the sound is set to OFF Step 9 This setting allows you to turn the strip expiration date indicator on or off This setting turns the funct...

Страница 10: ...trol Solution ranges are printed on the KetoSens test strip box Compare the result displayed on the meter to the KetoSens Control Solution range printed on the test strip box Before using a new meter...

Страница 11: ...toms are inconsistent with the blood Ketone test results and you feel that the meter or test strips are not working properly If you drop or damage the meter Note The KetoSens Control Solution can be p...

Страница 12: ...ion test by referring to the Notes on page 21 Discard the used control solution and repeat the test using a new bottle of control solution Caution To reduce the chance of infection All components that...

Страница 13: ...Device Step 5 The lancing device has six puncture depth settings numbered 0 through 5 0 for a shallow puncture 5 for a deeper puncture Choose a depth by rotating the top portion of the adjustable tip...

Страница 14: ...to form A minimum volume of 0 5 microliter is needed to fill the confirmation window actual size of 0 5 L Step 9 After the symbol appears on the screen apply the blood sample to the narrow end of the...

Страница 15: ...strips safely in a disposable container Blood ketone test result Step 1 Unscrew the lancing device tip Step 2 Place the protective cover on the lancet Push the lancet ejector forward with the thumb to...

Страница 16: ...Meter Memory The KetoSens meter can save up to 1 000 Ketone test results with time and date If the memory is full the oldest test result will be deleted and the latest test result will be stored Note...

Страница 17: ...a test strip press the and S buttons simultaneously for 3 seconds to enter the time alarm setting alarm 1 will be displayed while OFF blinks on the screen Step 2 On pressing the button alarm 1 is set...

Страница 18: ...e meter and lancing device should be disinfected prior to use by the second person Pre cleaning and Disinfection Pre cleaning procedure is needed to clean dirt as well as blood and other body fluids o...

Страница 19: ...tact the manufacturer or visit their website as listed above Name Manufacturer EPA registration number Active ingredients Clorox Healthcare Bleach Germicidal Wipes Clorox Professional Products Company...

Страница 20: ...n the meter or lancing device When a part of the segment on the meter display does not flash When control solution test results do not fall within the stated range on the test strip box Caution Do not...

Страница 21: ...sample was applied before the symbol appeared Repeat the test with a new test strip and wait until the symbol appears before applying the blood or control solution sample Message The blood sample is o...

Страница 22: ...ting a test strip The test does not start even after applying the blood sample on the test strip The test result does not match the way you feel General Troubleshooting Note If the problem is not reso...

Страница 23: ...avg 3 4 mmol L CV 3 7 Blood avg 5 2 mmol L CV 3 5 Blood avg 6 9 mmol L CV 3 6 Precision Precision studies were performed in a laboratory using the KetoSens Brand of BKM Systems Between Run Precision C...

Страница 24: ...ling tampering use or service of the meter Any claim must be made within the warranty period i SENS Inc will at its discretion repair or replace a defective meter or meter part that is covered by this...

Страница 25: ...www i sens com 49 MEMO...

Страница 26: ...50 www i sens com MEMO...

Страница 27: ...being achieved ask a doctor to review management Reference Blood Ketone Measurement UHL Guideline 2013 Lo 0 5 mmol L Re test for blood ketones only if the blood glucose levels are higher than 120 mg...

Страница 28: ...he back of this card for important information about ketone levels Keto Diet Ketone Blood Level Ranges May require medical attention Please check with your medical provider mmol L References http www...

Отзывы: