36
www.i-sens.com
www.i-sens.com
37
Step 4
On pressing the
◀
button, the number
indicating the minute will start blinking.
Press the
▶
button to set the minute.
Step 5
Press the
S
button to finish and to enter
the 'alarm 2' mode.
Repeat steps 2 to 5 to set the remaining
time alarms (alarm 2 and 3).
Step 6
Press and hold the
S
button for 3 seconds to finish and
turn the meter off.
Caring for Your System
• To minimize the risk of transmission of blood-borne
pathogens, the pre-cleaning and disinfection procedure
should be performed as recommended in the instructions
below.
• Wash your hands thoroughly with soap and water after
handling the meter, lancing device, or test strips.
• If the meter is being operated by a second person who
is providing testing assistance to the user, the meter and
lancing device should be disinfected prior to use by the
second person.
Pre-cleaning and Disinfection:
Pre-cleaning procedure is needed to clean dirt as well as
blood and other body fluids on the exterior of the meter and
lancing device before performing the disinfection procedure.
The disinfection procedure is needed to prevent
transmission of blood-borne pathogens.
• For the meter and lancing device, this pre-cleaning and
disinfection procedure should be performed
once per
week
.
Note:
The life span of a KetoSens meter is 5 years. We
recommend cleaning and disinfecting both the meter and
lancing device at least once per week. We have validated
a total of 260 cleaning and disinfecting cycles (260
Cleaning and 260 disinfection cycles) to represent weekly
cleaning and disinfecting over the use life of your meter
and lancing device.
1 pre-cleaning and 1 disinfection cycles per week * 52
weeks per year * 5 years = 260 pre-cleaning and 260
disinfection cycles