26
Blood Glucose Monitoring System
27
www.i-sens.com
Step 10
The test result will appear after the meter
counts down from 5 to 1. The result will
be automatically stored in the meter’s
memory. If the test strip is removed after
the test result is displayed, the meter will
automatically switch off after 3 seconds.
Discard used test strips safely in
disposable containers.
The meter may switch off if the blood sample is not applied
within 2 minutes of the
icon appearing on the screen.
If the meter turns off, remove the strip and reinsert after
switching on the meter.
Note
Obtain a blood sample using the lancing
device. Place the device against the pad
of the finger. The best puncture sites are
on the middle or ring fingers. Press the
release button. Remove the device from
the finger.
Wait a few seconds for a blood drop to form.
A minimum volume of 0.5 microliter is needed to fill the
DPOGJSNBUJPOXJOEPX BDUVBMTJ[FPG
㎕
: )
After the
symbol appears on the
screen, apply the blood sample to the
narrow end of the test strip till the meter
beeps. If the confirmation window is not
filled before the meter finishes counting
down then discard the test strip and
insert a new one.
If confirmation window is not filled in
time because of abnormal viscosity or
insufficient volume, Er4 message will
appear.
Step 9
Good
Sample
Insufficient
Sample
Applying Blood Sample
Step 8