28
www.i-sens.com
www.i-sens.com
29
Step 4
The display segments will rotate
clockwise and a test result will
appear after the meter counts
down from 5 to 1.
When flagged, the result is stored
in the meter's memory but it is
not included in the averages.
Step 5
Compare the result displayed on
the meter to the range printed
on the test strip vial. The result
should fall within the range.
Caution:
The range printed on the test strip vial is for the
CareSens Control Solution only. It has nothing to do with your
blood glucose level.
Control Solution Range
Control A:
101-151 mg/dL (5.6-8.4 mmol/L)
Control B:
184-276 mg/dL (10.2-15.3 mmol/L)
Note:
The CareSens Control Solution can be purchased separately.
Please contact your authorised i-SENS sales representative.
Step 3
Shake the bottle before
each test.
Remove the cap and
squeeze the bottle to
discard the first drop. Then
wipe the tip with a clean
tissue or cloth. Dispense a
drop of control solution onto a clean non-absorbent surface. It
helps to squeeze a drop onto the top of the cap as shown. After
the
symbol appears on the display, apply the solution
to the narrow edge of the test strip until the meter beeps. Make
sure the confirmation window fills completely.
Note:
The meter may switch off if the control solution sample
is not applied within 2 minutes of the
symbol
appearing on the screen. If the meter turns off, remove the
strip, reinsert, and start from step 1.