Step 11

This mode allows you to turn the strip expiration date indicator 

on or off. This mode turns the function on or off only. See page 

24 to set the strip expiration date.

When 'EP' appears on the screen, press the 

 or 

 button. The 

screen will display 'On' or 'OFF'. Press the 

S

 button to confirm 

the setting. 

If you do not want to set the indicator, press the 

S

 button while 

the screen displays 'OFF'.

Turning on the Strip Expiration Date Indicator

Note:

 

If the pre-set expiration date 

expires, the meter will display 

EP when the test strip is 

inserted. EP shows alternately 

also when the test result is 

displayed right after the test. 

If the expiration date is set to 

October of 2017, the meter 

will display EP at the start of 

November, 2017.

Step 10 

On pressing the 

 or 

 button, the 

screen will display 'On' or 'OFF'. Press the 

 button to confirm the selection.

The meter will beep in the following 

instances if set to On.

t

When you push a button to turn on the 

meter

t

When the test strip is inserted in the 

meter

t

When the blood sample is absorbed 

into the test strip and the test starts

t

When the test result is displayed

t

When you press and hold the 

◀ 

button to set the post-meal (PP2) alarm

t

When it is time for a pre-set blood 

glucose test

Setting the Sound On/OFF

Note:

 

The 

symbol is displayed only when the sound is set to 

OFF.

If the sound is set to OFF, none of the sound functions will work.

After setting the sound, press the 

button to progress to the next 

step.

Содержание CareSens N Premier

Страница 1: ...er license All other trademarks and trade names are those of their respective owners WARNING Turn off the Bluetooth feature in areas where the use of wireless devices is restricted such as hospitals s...

Страница 2: ...he Strip Expiration Date Indicator 25 Checking the System 26 Control Solution Testing 27 Comparing the Control Solution Test Results 30 Testing Using the Lancing Device 31 Preparing the Lancing Device...

Страница 3: ...ors in monitoring by using the no coding function t 5IF BSF4FOT 1SFNJFS CMPPE HMVDPTF NFUFS TIPVME CF used only with the CareSens N Test Strips t O BCOPSNBMMZ IJHI PS MPX SFE CMPPE DFMM DPVOU IFNBUPDS...

Страница 4: ...logy t Operating range distance maximum 10 meters unobstructed t Operating channels 40 channels t Security encryption 128 bit AES Advanced encryption standard Temperature 5 50 C 41 122 F Relative humi...

Страница 5: ...o place Caring for Your System Use a soft cloth or tissue to wipe the meter exterior If necessary dip the soft cloth or tissue in a small amount of alcohol Do not use organic solvents such as benzene...

Страница 6: ...U TUSJQ GPS UFTUJOH BOE use the strip immediately t BOEMF UFTU TUSJQT POMZ XJUI DMFBO BOE ESZ IBOET t P OPU CFOE DVU PS BMUFS UFTU TUSJQT JO BOZ XBZ t PS EFUBJMFE TUPSBHF BOE VTBHF JOGPSNBUJPO SFGFS U...

Страница 7: ...al test flag used for tests done after eating Fasting test flag used for tests done after fasting for at least 8 hours Month Day Hour Minute 10 12 7 8 9 11 13 14 15 1 2 3 5 4 6 7 11 10 9 15 12 13 14 N...

Страница 8: ...our smartphone The devices must be within 10 metres of each other to pair and synchronise Before pairing your meter and smartphone download and install the SmartLog mobile app on your smartphone Step...

Страница 9: ...splays FAIL and then OFF repeat steps 2 to 5 4 The meter will show six digit PIN number 5 Enter the PIN number into the SmartLog mobile app and touch OK Make sure the PIN you enter on your smartphone...

Страница 10: ...T OPU EJTQMBZFE JO UIF IPVS format After selecting the format press the S button to go to the next step Adjusting the Date and Time Step 4 Setting the Year Press the or button to adjust until the corr...

Страница 11: ...fter the test If the expiration date is set to October of 2017 the meter will display EP at the start of November 2017 Step 10 On pressing the or button the screen will display On or OFF Press the S b...

Страница 12: ...ng press and hold the S button for 3 seconds to turn off the meter Step 12 This mode allows you to turn the hypoglycemia indicator possible low blood sugar On or OFF and to select the desired level fo...

Страница 13: ...drops and wipe the tip clean t MPTF UIF DPOUSPM TPMVUJPO CPUUMF UJHIUMZ BOE TUPSF BU B temperature between 8 30 C 46 86 F Press and hold Checking the System You may check your meter and test strips us...

Страница 14: ...276 mg dL 10 2 15 3 mmol L Note The CareSens Control Solution can be purchased separately Please contact your authorised i SENS sales representative Step 3 Shake the bottle before each test Remove the...

Страница 15: ...Results The test result of each control solution should be within the range printed on the label of the test strip vial Repeat the control solution test if the test result falls outside of the range O...

Страница 16: ...vary for various people at different sample sites The lancing device s adjustable tip allows the best depth of skin penetration to get an adequate sample size Note 1 least penetration of lancet into...

Страница 17: ...and stickiness or insufficient volume the Er4 message may appear Good Sample Insufficient Sample Step 8 Obtain a blood sample using the lancing device Place the device against the pad of the finger T...

Страница 18: ...the meter must be turned On t The meter and a smartphone must be paired t The SmartLog mobile app must be launched The meter will transmit data in the following cases t When the strip is ejected afte...

Страница 19: ...ree of visible veins and hair and away from bones Gently massage the sample site to help blood circulation to minimise result differences between fingertip and alternate site sampling Firmly press and...

Страница 20: ...toms t G UIF TBNQMF ESPQ PG CMPPE SVOT PS TQSFBET EVF UP DPOUBDU with hair or with a line in your palm do not use that sample Try puncturing again in a smoother area Things to KnowWhen Using AST Pleas...

Страница 21: ...screendisplays Pc whenitisconnectedtothecomputerusingthedata cable Formoreinformation contactyourauthorised i SENSsalesrepresentativeorvisitwww i sens com Note Test results can also be transferred wir...

Страница 22: ...30 90days Fasting average 1 7 14 30 90days Meter Memory Step 1 Press the or S button to turn the meter on The current date and time will be displayed at the bottom of the screen followed by the 1 day...

Страница 23: ...m the most recent and ending with the oldest Press the button to return to the result seen previously After checking the stored test results hold the S button to turn off the meter Thenumberoftests wi...

Страница 24: ...e button for 3 seconds PP2 bell symbol and OFF will appear on the screen Then the screen will change automatically to the memory recall mode without bell symbol displayed Four types of alarms can be s...

Страница 25: ...larm1 OFF will blink on the screen Step 3 Pres the button to adjust the time of alarm1 A number representing the hour will blink on the screen Press the button to set the hour Setting the Time Alarms...

Страница 26: ...ithin the operating range 5 50 C and repeat the test after the meter and test strips have reached a temperature within the operating range This error message may appear when the wrong blood glucose te...

Страница 27: ...please contact your authorised i SENS sales representative Performance Characteristics The performance of CareSens N Premier Blood Glucose Monitoring System has been evaluated in laboratory and in cl...

Страница 28: ...t of glucose concentration Difference Averages NO Interferent 1 Acetaminophen 2 3 mg dL 0 1 mmo L 3 3 2 Ascorbic acid 7 3 mg dL 0 4 mmol L 0 9 3 Bilirubin unconjugated 0 1 mg dL 0 01 mmol L 1 4 4 Ceft...

Страница 29: ...he medical laboratory values at glucose concentrations below 100 mg dL 5 55 mmol L and 98 9 within 15 of the medical laboratory values at glucose concentrations at or above 100 mg dL 5 55 mmol L Warra...

Страница 30: ...60 www i sens com MEMO...

Отзывы: