Control Solution Testing
Step 1
Insert a test strip into the meter’s test strip port,
with the contact bars facing upwards.
Gently push the test strip into the port until the
meter beeps. Be careful not to bend the strip
while pushing it in.
The
symbol will show up.
Step 2
You can flag the control solution test result
by pressing and holding the
▼
button for 3
seconds. To undo the control solution flag, press
and hold the
▼
button for 3 seconds again.
You may do a control solution test:
• When you want to practice the test procedure using the
control solution instead of blood,
• When using the meter for the first time,
• Whenever you open a new vial of test strips,
• If the meter or test strips do not function properly,
• If your symptoms are inconsistent with the blood glucose
test results and you feel that the meter or test strips are not
working properly,
• If you drop or damage the meter.
Step 4
The display segments will rotate
clockwise and a test result will
appear after the meter counts
down from 5 to 1.
When flagged, the result is stored
in the meter's memory but it is not
included in the averages.
Note:
The meter may switch off if the control solution
sample is not applied within 2 minutes of the
symbol
appearing on the screen. If the meter turns off, remove the
strip, reinsert, and start from step 1.
Step 3
Shake the bottle before each test.
Remove the cap and squeeze the
bottle to discard the first drop.
Then wipe the tip with a clean
tissue or cloth. Dispense a drop
of control solution onto a clean
non-absorbent surface. It helps
to squeeze a drop onto the top of the cap as shown. After the
symbol appears on the screen, apply the solution to the
narrow edge of the test strip until the meter beeps. Make sure
the confirmation window fills completely.