I-MED I-PEN Скачать руководство пользователя страница 8

14

15

9. ELECTROMAGNETIC EMISSIONS

10. ELECTROMAGNETIC IMMUNITY

NOTES

 

   

The I-PEN

®

 requires special precautions with regard to electromagnetic compatibility. 

    

It must be installed and prepared for use as described in Section 4. Performing an Osmolarity 

Measurement.

   

Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere 

with the I-PEN

®

   

The recommended separation distances in this paragraph must therefore be complied with.

   

The I-PEN

®

 must not be used near or on top of another device. If this cannot be avoided, it is necessary 

– before clinical use – to check the equipment for correct operation under the conditions of use. 

   

The use of accessories other than those specified or sold by I- MED Pharma Inc. as replacement parts 

may have the consequence of increasing the emissions or decreasing the immunity of the unit. 

    

I-PEN

®

 is intended for use in the electromagnetic environment specified in the following tables. 

This is not a life- sustaining device. 

•  

  

The user and/or installer of the unit must ensure that it is used in such an environment.

Guidance and Manufacturer’s Declaration -  Electromagnetic Emissions

The I-PEN

®

 is intended for use in the electromagnetic environment specified below. The customer or the 

user of the I-PEN

®

 device should assure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment -  

Guidance

RF emissions

Test: CISPR 11

Group 1

The I-PEN

®

 uses RF energy only for 

its internal function. Therefore, its RF 

emissions are very low and are not likely 

to cause any interference in nearby 

electronic equipment.

RF emissions 

CISPR 11

Class B

Harmonic emissions

IEC  61000- 3- 2

Not applicable

Voltage fluctuations/flicker 

emissions IEC 61000- 3-3

Not applicable

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

The I-PEN

®

 is intended for use in the electromagnetic environment specified below. The customer or the 

user of the I-PEN

®

 device should assure that it is used in such an environment.

Immunity Test

IEC 60601-1-2 Test 

Level

Compliance 

Level

Electromagnetic 

Environment- Guidance

Electrostatic 

discharge (ESD)

IEC 61000-4-2

±6 kV contact 

±8 kV air

±8 kV contact

±8 kV air

Floors should be wood, 

concrete or ceramic tile. 

If floors are covered with 

synthetic material, the 

relative humidity should be 

at least 30%.

Electrical fast

transient/burst

IEC 61000-4-4

±2 kV

for power supply lines 

±1 kV

for input/ Output lines

Not applicable

Surge 

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

Not applicable

Voltage fluctuations/

flicker emissions IEC 

61000- 3-3

<5 %UT

(>95 %dip in UT)

for 0.5 cycle

40 %UT

(60 %dip in UT)

for 5 cycles <5 %UT

70 %UT

(30 %dip in UT)

for 25 cycles

<5 %UT <5 %UT

(>95 %dip in UT)

for 5 s

Not applicable

Power frequency 

(50/60 Hz) magnetic 

field IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic 

fields should be at levels 

characteristic of a typical 

public low-voltage power 

supply network that supplies 

buildings used for domestic 

purposes, commercial or 

hospital, clinic environment.

I-PEN

®

 User Guide  

|

  ELECTROMAGNETIC EMISSIONS

I-PEN

®

 User Guide  

|

  ELECTROMAGNETIC IMMUNITY

Содержание I-PEN

Страница 1: ...FOR QUANTITATIVE MEASUREMENT OF OSMOLARITY OF OCULAR TISSUES Osmolarity System USER MANUAL osdcare com...

Страница 2: ...NDITIONS 13 8 TECHNICAL SPECIFICATIONS 13 9 ELECTROMAGNETIC EMISSIONS 14 10 ELECTROMAGNETIC IMMUNITY 15 10 1 Recommended Separation Distances 17 10 1 1 Applicable Standards 17 11 LABELS AND SYMBOLS 18...

Страница 3: ...smolarity Single Use Sensor in conjunction with the I PEN Osmolarity System provides a quick and simple method for determining tear osmolarity using impedance measurements of the saline concentration...

Страница 4: ...WARNING Do not test patients who have used eye drops within two hoursprior to testing WARNING Do not test patients wearing makeup on eyelids WARNING Do not test patients within 10 minutes after remova...

Страница 5: ...n 5 insert the disposable Single Use Sensor CAUTION Replace the device if a beep is not heard after turning it on CAUTION It is important to visually inspect the Single Use Sensor before use In the ca...

Страница 6: ...ontact with the conjunctiva 6 The I PEN will take make an audible beep after several seconds and display the reading on the LCD screen 4 5 TAKING A READING 1 Ask the patient to gently squeeze their ey...

Страница 7: ...sage does not display Battery not installed Device malfunction Verify that the correct type of battery is installed and that the battery is fresh Contact I PEN customer support A beep is not heard whe...

Страница 8: ...s RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Harmonic em...

Страница 9: ...be necessary such as re orienting or relocating the I PEN 10 1 RECOMMENDED SEPARATION DISTANCES 10 1 1 APPLICABLE STANDARDS Recommended separation distances between portable and mobile RF communicati...

Страница 10: ...requirements and standards and are briefly reviewed below REF 600 SN nn xxxx mm yyyy LifeCare Ltd 2 Zipori Street Tiberias 1424602 Israel PRT 14 200 Rev 01 CAUTION At the end of its useful life the sy...

Страница 11: ...I MED Pharma Inc 1601 St Regis Blvd Dollard des Ormeaux QC Canada H9B 3H7 Tel 514 685 8118 Toll free 800 463 1008 Fax 514 685 8998 info imedpharma com SEE THE DIFFERENCE Product IPUMEN 0718...

Отзывы: