6. CLEANING AND MAINTENANCE

7. OPERATING AND STORAGE CONDITIONS

6.1. CLEANING

6.2. MAINTENANCE

6.1.1. I-PEN

®

 DEVICE

6.1.2. SINGLE-USE-SENSORS

6.2.1. TROUBLESHOOTING

The I-PEN

®

 can be cleaned with a damp cloth or alcohol wipe as required. When cleaning, it is 

important to keep the electronic contacts of the control unit and Reader dry. The electronic contacts 

and docking port should also be kept free of dust and dirt.

To ensure warranty coverage and reliable system operation, defective system components should  

be serviced and/or replaced exclusively by I- MED Pharma Inc. authorized personnel, and replacement 

parts should be those specified by the manufacturer. It is important to use and store the device within 

the environmental conditions shown in the table below.

Single-Use-Sensors are for use on a single eye. Never reuse or try to clean a Single-Use-Sensor.

Single-Use-Sensors are for single use only. Never reuse or try to clean a Single-Use-Sensor.

The I-PEN

®

 Osmolarity System is designed to work without direct service or preventive maintenance. 

If quality checks fail, contact I-PEN

®

 Customer Support.

Battery should be replaced when “Low Bat” indication appears on the screen.

Condition

Possible Cause

Recommended Action

The “I-PEN Ready” message

does not display

Battery not installed.

Device malfunction.

Verify that the correct type of battery is 

installed, and that the battery is fresh.

Contact I-PEN

®

 customer support.

A “beep” is not heard when

the device is turned on

Device malfunction.

Contact I-PEN

®

 customer support.

“Low Bat” indication appears 

on screen

Battery close to

end-of-life.

Replace battery.

Screen goes dim

Battery close to

end-of-life.

Replace battery.

Transport and Storage Temperature

2°–35°C/36°–95°F

Transport and Storage Relative Humidity

10–85%  non- condensing

Transport and Storage Altitude

0–2,000 meters

Operating Temperature

15°– 30°C/59–86°F

Operating Altitude

0–2,000 meters

Operating Relative Humidity

10–85%  non- condensing

NOTE: 

If the Recommended Action does not solve the problem, 

contact I- MED Pharma Inc. Customer Support.

Single-Use-Sensors may be ordered on line at www.imedpharma.com or by calling your 

representative at I- MED Pharma Inc. Tel. Number: (800)  463- 1008 or (514)  685- 8118.

I-PEN

®

 User Guide  

|

  CLEANING AND MAINTENANCE

I-PEN

®

 User Guide  

|

  OPERATING AND STORAGE CONDITIONS  |  TECHNICAL SPECIFICATIONS

8. TECHNICAL SPECIFICATIONS

Calibration

No calibration required

Degree of Protection Against Electric Shock

BF Type applied part

Size (not including probe holder)

W 140mm

L 223mm

H 140mm

Weight

50 gm

Battery

CR2032

Frequency

80 Hz

Peak Voltage

± 1.5V

Current Source

Max100 μA AC

Sinus Distortions

5%

WARNING: 

Cleaning fluids should never be used on the Single-Use-Sensor.

Содержание I-PEN

Страница 1: ...FOR QUANTITATIVE MEASUREMENT OF OSMOLARITY OF OCULAR TISSUES Osmolarity System USER MANUAL osdcare com...

Страница 2: ...NDITIONS 13 8 TECHNICAL SPECIFICATIONS 13 9 ELECTROMAGNETIC EMISSIONS 14 10 ELECTROMAGNETIC IMMUNITY 15 10 1 Recommended Separation Distances 17 10 1 1 Applicable Standards 17 11 LABELS AND SYMBOLS 18...

Страница 3: ...smolarity Single Use Sensor in conjunction with the I PEN Osmolarity System provides a quick and simple method for determining tear osmolarity using impedance measurements of the saline concentration...

Страница 4: ...WARNING Do not test patients who have used eye drops within two hoursprior to testing WARNING Do not test patients wearing makeup on eyelids WARNING Do not test patients within 10 minutes after remova...

Страница 5: ...n 5 insert the disposable Single Use Sensor CAUTION Replace the device if a beep is not heard after turning it on CAUTION It is important to visually inspect the Single Use Sensor before use In the ca...

Страница 6: ...ontact with the conjunctiva 6 The I PEN will take make an audible beep after several seconds and display the reading on the LCD screen 4 5 TAKING A READING 1 Ask the patient to gently squeeze their ey...

Страница 7: ...sage does not display Battery not installed Device malfunction Verify that the correct type of battery is installed and that the battery is fresh Contact I PEN customer support A beep is not heard whe...

Страница 8: ...s RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Harmonic em...

Страница 9: ...be necessary such as re orienting or relocating the I PEN 10 1 RECOMMENDED SEPARATION DISTANCES 10 1 1 APPLICABLE STANDARDS Recommended separation distances between portable and mobile RF communicati...

Страница 10: ...requirements and standards and are briefly reviewed below REF 600 SN nn xxxx mm yyyy LifeCare Ltd 2 Zipori Street Tiberias 1424602 Israel PRT 14 200 Rev 01 CAUTION At the end of its useful life the sy...

Страница 11: ...I MED Pharma Inc 1601 St Regis Blvd Dollard des Ormeaux QC Canada H9B 3H7 Tel 514 685 8118 Toll free 800 463 1008 Fax 514 685 8998 info imedpharma com SEE THE DIFFERENCE Product IPUMEN 0718...

Отзывы: