User Guide
21
INS2051 Rev A
19 Press the SELECT/CONTINUE push
button until the arrow
is next to the
Hi Temp Cal Pt
option.
Press the SCROLL and SELECT
buttons at the same time to select the
Hi Temp Cal Pt
option.
20 The LCD display changes to a screen
that compares the Displayed and
Measured temperatures.
If needed, use the SCROLL button to
increase the value of the Measured
temperature or the
SELECT/CONTINUE button to
decrease the value of the Measured
temperature to the value displayed on
the thermometer.
Note: When entering the Measured
temperature, do not include decimal
places.
21 Press the SCROLL and
SELECT/CONTINUE buttons at the
same time to set the Measured
temperature.
If necessary, the thermal block
automatically adjusts the Plate
Temperature to align with the Set Point.
22 Calibration is complete. Turn off power
to the thermal block by toggling the
on/off switch on the back panel.
Lo Temp Cal Pt
Hi Temp Cal Pt
Displayed: 95C
Measured: 96C
Set Point: 95C
Plate Temp: 95C
Exit
Calibrate