User Guide
11
INS2051 Rev A
5
The LCD display changes to “37C 24E
Lysis”. A timer in the bottom right corner
counts down from 30:00 and shows the
number of minutes remaining in this
step.
6
After lysis at 37 °C is complete, the blue
ACTION NEEDED LED activates and
the thermal block sounds two audible
beeps to signal that user action is
required. The LCD display changes to
“Remove & Press Continue”.
7
Remove the rack of cluster tubes from
the metal tube holder.
Press the SELECT/CONTINUE button.
8
The unit beeps once and the LCD
display changes as the thermal block
begins equilibrating to 55 °C.
While the block is heating, follow the
instructions in the BAX System User
Guide to add additional lysing reagents
to each sample.
Note: Do not re-load samples into the
thermal block until the re-load prompt
appears on the LCD display.
9
When the thermal block has reached
55 °C, the unit sounds two audible
beeps and the blue ACTION NEEDED
LED activates. The LCD display
changes to “Reload Samples”.
10 After the “Reload Samples” prompt
appears, place the rack of samples with
additional lysing reagents into the metal
tube holder.
Press the SELECT/CONTINUE button.
37C 24E Lysis
24E 29:27
Remove & Press
Cont.
Equilibr.@ 55C
24E
ReLoad Samples
24E