18
LUMA X
3/2022
4.2
Firmware upgrade and factory reset
In order to upgrade the firmware of LUMA X or carry out a factory-reset, LUMA X must be put
into
Bootloader
mode. The following steps will upgrade LUMA X firmware and/or carry out a
factory reset:
Start
LumaConf
Ensure that the LUMA X is connected to a computer through the serial connection. This
is most easily achieved by using Hydromea’s power/data cable-box which provides a
serial connection through a USB-to-serial conversion within the power/data cable and
can be accessed via a micro-USB port.
Ensure that the LUMA X is connected to a power supply
In
LumaConf
set
Interface
to the correct serial port
Click on
Firmware update
within
LumaConf.
This will open the
Bootloader
dialog box
(see Figure 11).
Cycle the power for LUMA X
The user now has a few seconds after start-up to discover the LUMA X by clicking the
Discover
button. Once it appears in the
Devices
window, the LUMA X will stay in
bootloader mode and more options will be available to the user.
o
Firmware factory reset:
select the proper firmware file by clicking on
Select …
and then start the update by clicking on
Write Flash
. Wait for the
progress bar to reach 100%. After a firmware update a factory reset will also be
carried out automatically.
Note: please make sure that you select the proper firmware image, the last
digits of the filename LX########## need to match the serial number
engraved on the modem.
o
Factory reset only:
clicking on
Clear Settings
will factory reset all LUMA X
settings.
Clicking on
start Application
will cause the attached LUMA X to exit the bootloader and
restart. The LUMA X will restart in
config
mode as a firmware update will also cause a
factory reset which sets the
Default state
after reboot to
config
.
Closing the bootloader window and selecting Ethernet in
Interface
will then lead to the
LUMA X showing up in
LumaConf
.
Note: a limited set of options is available in the
Bootloader
menu when one or multiple
LUMA X is/are connected via Ethernet and in
config
mode. In this case the
Bootloader
window will list all LUMA X connected to the network with their last IP byte as an index.
The only available actions in this mode are reading the unique serial number of a LUMA X
by clicking on
Get Info
which then shows the serial number in
Processor ID
or to soft reset
the attached modem(s) by clicking on
Reset.