P/N 20-07665-00 Rev
L
© Hydro Systems Company, Inc. 201
8
HydroSystemsCo.com Toll Free: 1.800.543.7184 11
Dead Unit (No Display) Troubleshooting
1. Verify Main Power (between 90 VAC and 249 VAC, 50/60 Hz) is present on Main Power Cable (or
Terminals) and Power Switch (located behind cover in upper left corner) is On.
Yes? Go to step 2.
No? Restore Main Power source.
2. With a volt meter capable of reading AC voltages between 90VAC and 249 VAC, and DC voltages
around 20 VDC to 30 VDC, perform the following measurements on the Power Supply PCB:
2a. Recheck Main Power source, if Main Power present, replace Logic PCB.
Troubleshooting
Problem
Possible Cause
Solution
No voltage to main power terminals
1. Tripped breakers or blown fuses at
power source
1. Reset breakers or replace fuse at
power source
2. Main power wires
2. Check connections at plug or washer
power terminals
Voltage at main power terminals, but LED display
is not lit (see additional troubleshooting info above)
1. Power supply failure
1. Replace power supply
2. Damaged logic circuit board
2. Replace logic circuit board
Pumps run slowly or not at all
1. Pumps not plugged in properly
1. Verify pump electrical connectors are
plugged in properly
2. Pump failure
2. Replace failed pump
Password will not access program mode
1. Password has been changed
1. Call Hydro Systems Customer Service for
new Logic PCB
Pump runs but will not deliver product
1. Squeeze tube failure
1. Replace squeeze tube
2. Product tubing plugged
2. Replace product tubing
3. Pump roller failure
3. Replace pump roller
Poor results
1. Squeeze tube failure
1. Replace squeeze tube
2. Product tubing plugged
2. Replace product tubing
3. Pump roller failure
3. Replace pump roller
Dispenser not dispensing upon request from washer
1. Incorrect Operating Mode (Op)
selected from programming menu
1. Select appropriate Operating Mode under
(Op) in the programming menu
2. Signal wires are loose or
wired incorrectly
2. Check signal wiring to washer