FIGURE 3 (Sampling Examples)
Sample
Pump
Low Pressure
Sample Water
Source
Flush Valve
to Drain
Pressure entering
RAH-210 must be
reduced to 5 psig
(0.3 bar) or less
Sample Water
Flow Meter
Y-Strainer
Pressure Reducing
Valve Assembly
Pressure Reducing
Valve Assembly
Grab
Sample
Valve
To Drain
Pressurized
Sample Water
Source
It should also be considered that the sample point should be located such that the residual
reading can be used as a control signal for the chemical injection. Especially, it should be
considered that if there is a long time delay between chemical injection changes and the
change being detected by the measurement cell, then chemical injection control is adversely
aff ected. The delay time should be kept as short as possible. We recommend that the time
be less than 5 minutes.
8