HDA 5500 for HLB 1000 and CS 1000
- 19 -
9.3 Sequential analogue output of the HLB 1000
The analogue output signal of the measured values is output at PIN 4 of the HLB 1000 to the
HDA 5500 sequentially, as follows:
t
Start
Rel.
Visco
Rel.
DC
Rel.
humidity
Temp.
Status
Start
20mA
4mA
I
2s
30s
2s
Explanation: Output of sequential analogue output:
(each 4..20 mA)
O
utput signal
Duration
Start
signal:
20
mA
2
s
Signal 1:
Relative change in the viscosity
-30% ... +30%
2 s
Pause (4 mA)
2 s
Signal 2:
Relative change in the dielectric constant
-30% ... +30%
2 s
Pause (4 mA)
2 s
Signal 3:
Relative humidity
0% ... +100%
2 s
Pause (4 mA)
2 s
Signal
4: Temperature
-25°C...
+100°C
2
s
Pause (4 mA)
2 s
Signal 5:
Status signal
For levels see table below
2 s
Pause before the next cycle:
4 mA
30 s
Status signal
Status display HDA 5500
Status
Level of the status signal
0
Reference phase (no error)
5 mA
1
Function phase (no error)
7 mA
2 or 3
Fluid condition out of range
9 or 11 mA
4 to 7
Internal error
13 to 19 mA
Note:
The exact description of the status signal can be found in the Manual for the HLB 1000.
Due to possible tolerances, we recommend setting the threshold approx. 0.25 mA below the ideal value
given in the table.
Содержание HDA 5500 SERIES
Страница 2: ...HDA 5500 für HLB1000 und CS1000 ...
Страница 30: ...HDA 5500 für HLB1000 und CS1000 29 15 Geräteabmessungen ...
Страница 60: ...HDA 5500 for HLB 1000 and CS 1000 29 15 Unit dimensions Control panel thickness Control panel cut out ...
Страница 90: ...HDA 5500 pour HLB 1000 et CS 1000 29 15 Dimensions épaisseur paroi Découpe ...