AS 1000 - Reading out measured values
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 29 / 36
BeWa CSI-B-1 3424452b 2012-04-20 en-us.doc
2012-04-20
AS 1000 - Reading out measured values
If an AS 1000 is detected at the sensor input, then the CSI-B-1 outputs the following
analog signals:
Output
Value
OUT1
Saturation level S (0 ... 100 %)
Voltage U =
Voltage
U =
2 + 8 x S / 100%
Current I =
Current
I =
4 + 16 x S / 100%
S
=
(voltage U
−
2) × 100 / 8
S
=
(current I
−
4) × 100 / 16
OUT2
Temperature T (
−
25 … +100°C)
Voltage U =
Voltage
U =
2 + 8 x (T + 25°C) / 125°C
Current I =
Current
I =
4 + 16 x (T + 25°C) / 125°C
Temperature
=
(voltage U
−
2) × 125 / 8
Temperature
=
(current I
−
4) × 125 / 16