Reading the analog output
ISO + T
The ISO+T signal consists of 4 measured values which are transmitted time-
coded with the following time slices:
_
Ti
m
e
Signal
Variable
Signal duration
per pulse
in ms
Current (I) / Voltage (U)
1
Identifier
>2µm
300
High / Low
2
Measured
value
>2µm
3000
Current/Voltage for measured value
3
Identifier
>5µm
300
High / Low / High / Low
4
Measured
value
>5µm
3000
Current/Voltage for measured value
5
Identifier
>15µm
300
High / Low / High / Low / High / Low
6
Measured
value
>15µm
3000
Current/Voltage for measured value
7
Identifier
Temperature
300
High / Low / High / Low / High / Low / High / Low /
High / Low
8
Measured
value
Temperature
3000
Current/Voltage for measured value
4,0
2,0
4,1
2,05
I (mA)
U (V)
time (ms)
4,8
2,4
19,2
High
High
Low
Low
9,6
300
3000
3000
3000
1
2
4
6
8
3
5
7
19,5
9,85
4,3
2,15
19,7
9,75
0,0
19,8
9,9
ContaminationSensor CS1000
en(us)
Page 88 / 136
BeWa CS1000 3764916d 300 en-us 2017-09-25.docx
2017-09-25